The SafEye Quasar 900 Open-Path Gas Detection System is designed for the simultaneous detection of C1-C8 flammable gases. The system operates using a microprocessor-based flash source and detector unit, utilizing a spectral response of 2.0 - 3.0 microns to identify gas concentrations. It features a line-of-sight field of view with an alignment tolerance of ± 0.5° and includes heated optics to prevent condensation.

What's inside

  • Scope
  • Technical Description
  • Operating Modes
  • Technical Specifications
  • Installation Instructions
  • Operating Instructions
  • Maintenance Instructions
  • Troubleshooting

Specifications

Detected GasesSimultaneous detection of C1-C8 flammable gases
Response Time3 sec to T90
Spectral Response2.0 - 3.0 micron
Operating Voltage18-32 VDC
Water and Dust TightIP66 and IP68
Temperature Range-67°F (-55°C) to +149°F (+65°C)

Frequently asked questions

What is the maximum detection distance for the Quasar 900?

The maximum detection distance ranges from 23 ft (7 m) for the 901 model up to 660 ft (200 m) for the 904 model.

Is the Quasar 900 explosion-proof?

Yes, it is certified as ATEX & IECEx EX II 2 GD, Ex d e ib [ib Gb] IIB+H2 T4 Gb Ex tb IIIC T135°C Db.