Changhui SWP-TU Intelligent Vortex Flowmeter User Manual

SWP series intelligent instrument Intelligent vortex flowmeter adopts Karman vortex street theory, uses piezoelectric sensor to detect the frequency f, and then sends it to the single-chip microcomputer of the instrument form for calculation, processing and display after electronic circuit detection.


Changhui SWP-TU Intelligent Vortex Flowmeter User Manual - cover page
Brand
Changhui
Category
Other
Document type
User Manual
Language
Chinese
Pages
26
File format
PDF
File size
1.2 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
5DBCD1166DB953384DA42A40FA0CCBE2

About this manual

The Changhui SWP-TU series intelligent vortex flowmeter is designed to measure the volume flow of liquids, gases, and steam. It operates based on the Karman vortex street theory, utilizing a piezoelectric sensor to detect vortex frequency and a microcontroller for processing and display. The device is suitable for applications in the energy, water, petrochemical, and food industries. This manual is written in Chinese.

What's inside

  • Safety Precautions
  • Product Overview
  • Working Principle
  • Technical Specifications
  • Assembly and Installation
  • Electrical Connection

Specifications

Measurable FluidsLiquid, Gas, Steam
Nominal DiameterDN25, 40, 50, 80, 100, 150, 200
Pressure RatingPN2.5MPa
Process Temperature (Normal)-40 ~ +160℃
Process Temperature (High Temp)-40 ~ +280℃
Measurement Accuracy (Liquid)±1%
Measurement Accuracy (Gas)±1%
Measurement Accuracy (Steam)±1.5%
Protection RatingIP65
Operating Temperature-20 ℃~+55℃

Frequently asked questions

What fluids can the SWP-TU flowmeter measure?

It can measure the volume flow of liquids, gases, and steam.

What is the measurement accuracy for liquids and gases?

The accuracy is ±1% of the reading for both liquids and gases.

What is the minimum straight pipe length required upstream?

A minimum of 10D (10 times the pipe diameter) is required upstream.

What is the pressure rating of the sensor?

The sensor has a pressure rating of PN2.5MPa.