Technical requirements for medical protective masks
This Chinese national standard (GB 19083-2010) replaces the 2003 version and establishes mandatory requirements for self-filtering medical protective masks. These masks are designed for use in medical work environments to filter airborne particles and block droplets, blood, body fluids, and secretions.
The standard covers basic mask requirements such as coverage of the mouth and nose, facial fit, and the presence of a nose clip. It details specific performance criteria including filtering efficiency for non-oily particles, exhalation resistance, synthetic blood penetration resistance, surface wettability, microbial indicators, ethylene oxide residue limits, flammability, skin irritation, and fit factor. The document also provides detailed test methods for these parameters and requirements for labeling and packaging. This manual is written in Chinese.
| Exhalation Resistance | ≤343.2 Pa |
|---|---|
| Synthetic Blood Penetration | No penetration |
| Ethylene Oxide Residue | ≤10 µg/g |
| Flammability | Reburn time ≤5 s |
| Primary Skin Irritation Score | ≤1 |
| Fit Factor | ≥100 |
It applies to self-filtering medical protective masks used in medical work environments to filter airborne particles and block droplets, blood, body fluids, and secretions.
For masks sterilized by ethylene oxide, the residue must not exceed 10 µg/g.
The total fit factor of the mask should be no lower than 100.