This document is an instruction manual for two automatic steam sterilizers, the AUTESTER ST DRY PV-II 30 4002514 and the AUTESTER ST DRY PV-II 50 4002515, manufactured by JP SELECTA s.a.
The J.P. Selecta AUTESTER ST DRY PV-II 30 (4002514) and AUTESTER ST DRY PV-II 50 (4002515) are automatic autoclaves for steam sterilization with a pre-vacuum system. This instruction manual, code 80206.00 revision C dated July 2009, is provided in Spanish and English and covers safety, technical specifications, equipment description, installation, operation, configuration menus, test programs, maintenance and spare parts.
The manual describes the control panel, pre-established cycles, vacuum and Bowie-Dick test programs, and safety devices including a safety pressurestat adjusted to 2.35 bar and a safety valve adjusted to 2.5 bar. It also lists technical data such as 30 L and 50 L autoclave volumes, 230 V mains voltage, water quality requirements and RS-232 connection settings.
| Autoclave volume | 30 L (4002514) / 50 L (4002515) |
|---|---|
| Mains voltage | 230 V |
| Electrical power | 2800 W (30 L) / 3000 W (50 L) |
| Maximum current | 12.5 A (30 L) / 14 A (50 L) |
| Net weight | 90 kg (30 L) / 108 kg (50 L) |
| Useful chamber dimensions | Ø30 x 40 cm (30 L) / Ø30 x 70 cm (50 L) |
| Maximum metal load | 8 kg (30 L) / 12 kg (50 L) |
| Water volume per cycle | 1 L |
| Water quality | Demineralized, 15 to 200 microS/cm; pH 5 to 7 |
| Noise level | Below 70 dBA |
| Safety pressurestat | Adjusted to 2.35 bar |
| Safety valve | Adjusted to 2.5 bar |
It requires demineralized water between 15 and 200 microS/cm, with a pH between 5 and 7.
The door is blocked when the temperature is above 98 °C.
The safety valve is adjusted to 2.5 bar.
The autoclaves have a one-year warranty.