The EIP Labs Model 101A Spectrum Analyzer is a wide-frequency-range laboratory instrument covering 1-12.4 GHz, with usable coverage from 0.7 to 15.7 GHz in one band. It can plug into a Tektronix oscilloscope that accepts letter-series plug-ins or into the EIP Labs Model 201A Power Supply for use with another display.

This manual covers characteristics, operation, maintenance, replaceable parts, and diagrams. It includes front-panel controls, operating procedures for frequency- and time-domain measurements, calibration, and parts information.

What's inside

  • Characteristics
  • Description
  • Specifications
  • Operation
  • Controls, Connectors & Indicators
  • Operating Instructions
  • Principles of Operation
  • Maintenance
  • Replaceable Parts List
  • Diagrams

Specifications

Frequency coverage1-12.4 GHz (usable 0.7-15.7 GHz)
Dial frequency accuracy10 MHz + 3%
Frequency response+2 dB typical, ±4 dB maximum
Calibrated dispersion0-11.4 GHz (usable 0-15.0 GHz)
Dispersion resettabilityTypical 10 MHz
Dispersion dial accuracyTypical 10 MHz or ±0.5%, whichever is greater
Preselection3 resonator YIG filter, nominal 3 dB bandwidth 15 MHz
ResolutionTypical 15 MHz
Image rejectionGenerates no image signals
Intermod rejectionReduced by preselector selectivity over conventional microwave diode response
Type of receiverCrystal linear or log video with sweeping YIG preselector
Sensitivity (1 kHz bandwidth)-45 dBm minimum
Log accuracy±1 dB typical
Selectable video bandwidths1 kHz, 100 kHz, 3 MHz
Variable vertical log scaleVariable and calibratable, 2-20 dB/cm
Log video output voltage0.02 V/dB typical
Sweep rate1, 10, 30 Hz
Required associated equipmentScope or other analog display
Plug-ins available forTektronix oscilloscopes accepting letter-series plug-ins

Frequently asked questions

What frequency range does the Model 101A cover?

The Model 101A covers 1-12.4 GHz, with usable coverage from 0.7 to 15.7 GHz.

Can the Model 101A be used with a Tektronix oscilloscope?

Yes, it plugs into any Tektronix oscilloscope that accepts letter-series plug-ins, or into the EIP Labs Model 201A Power Supply.

How is the vertical display calibrated in dB per cm?

Set the vert display switch to log cal and adjust vert gain until the square-wave signal is exactly 1 em vertical; the display is then calibrated at 10 dB/em.

What video bandwidths are selectable?

The selectable video bandwidths are 1 kHz, 100 kHz, and 3 MHz.