Mettler Toledo Karl Fischer Titration Sampling Guide

This document describes the correct sampling techniques for Karl Fischer moisture determination.


Mettler Toledo Karl Fischer Titration Sampling Guide - cover page
Brand
METTLER-TOLEDO
Category
Other
Document type
Product Leaflet
Language
English
Pages
20
File format
PDF
File size
2.2 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
5E3DD1875C8E93F4755099160172F09A

About this manual

This brochure explains important rules and guidelines for drawing samples to determine water content according to Karl Fischer titration. It emphasizes avoiding water absorption from atmospheric moisture during sampling and storage to ensure accurate and reliable results. The guide incorporates the experience of Mettler Toledo's application chemists and covers various sampling techniques developed since 1935.

What's inside

  • Taking the sample
  • Storing the sample
  • Amount of sample
  • Liquid samples
  • Solid samples
  • Influence of Atmospheric Humidity
  • Titration stand
  • The drift
  • Good Measuring Practices
  • Literature and Guides

Specifications

Optimum water amount for coulometric analysis0.5 to 2 mg water per sample
Optimum water amount for volumetric titrationsapproximately 10 mg per sample
Acceptable relative standard deviation for values below 1000 ppm1 to 5%

Frequently asked questions

Why must samples be taken quickly?

Samples must be taken quickly to exclude, or at least minimize, the absorption or release of moisture which would change the water content.

Why are glass bottles preferable to plastic for storing samples?

Glass bottles are preferable because plastic is not completely gas-tight, allowing air moisture to penetrate and be absorbed by the sample.

What is the recommended titrant consumption for high accuracy in volumetric KF titration?

For high accuracy, the sample should require a titrant consumption between 30 and 70% of the nominal burette volume.

Download PDF (2.2 MB)