This document describes the correct sampling techniques for Karl Fischer moisture determination.
This brochure explains important rules and guidelines for drawing samples to determine water content according to Karl Fischer titration. It emphasizes avoiding water absorption from atmospheric moisture during sampling and storage to ensure accurate and reliable results. The guide incorporates the experience of Mettler Toledo's application chemists and covers various sampling techniques developed since 1935.
| Optimum water amount for coulometric analysis | 0.5 to 2 mg water per sample |
|---|---|
| Optimum water amount for volumetric titrations | approximately 10 mg per sample |
| Acceptable relative standard deviation for values below 1000 ppm | 1 to 5% |
Samples must be taken quickly to exclude, or at least minimize, the absorption or release of moisture which would change the water content.
Glass bottles are preferable because plastic is not completely gas-tight, allowing air moisture to penetrate and be absorbed by the sample.
For high accuracy, the sample should require a titrant consumption between 30 and 70% of the nominal burette volume.