B&C Electronics C 3630 2-Wire E.C. Transmitter Instruction Manual

The C 3630 is a digital 2-wire transmitter with a DIN rail housing. It can measure conductivity and temperature.


B&C Electronics C 3630 2-Wire E.C. Transmitter Instruction Manual - cover page
Brand
B&C Electronics
Category
Other Electronics
Document type
User Manual
Language
English
Pages
17
File format
PDF
File size
759 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
9F946F692EB81554CBA4E77BC8CBB3A5

About this manual

The B&C Electronics C 3630 is a 2-wire E.C. transmitter in a DIN rail housing. It works with 2- or 4-electrode E.C. cells and an RTD Pt100 temperature sensor, and provides an isolated 4/20 mA output proportional to conductivity. The manual covers physical and electrical installation, system checking, operation and calibration.

The manual also describes front-panel controls and terminal connections, scale, frequency and decimal-point selection, manual or automatic temperature compensation, cell constant adaptation, chemical calibration, preventive maintenance and troubleshooting. Specifications include conductivity scales from 0/199.9 µS to 0/19.99 mS, a -10.0/+120.0 °C temperature scale and 10/30 Vdc power.

What's inside

  • Functional description
  • Physical description
  • Specifications
  • Physical installation
  • Electrical installation
  • System checking
  • Operating the system
  • Normal operation
  • Preventive maintenance
  • Troubleshooting guide

Specifications

DisplayLCD
Input2-electrodes E.C. cell; 4-electrodes E.C. cell; RTD Pt100 2 or 3 wire
Output4/20 mA isolated
Conductivity scales0/199.9 µS, 0/1999 µS, 0/19.99 mS
Temperature scale-10.0/+120.0 °C
Temperature compensationManual or automatic
Power supply10/30 Vdc
Operating temperature0/50 °C
Dimensions105 x 95 x 58 mm (6 modules)
Net weight200 g
MountingDIN Rail mountable

Frequently asked questions

What power supply does the C 3630 require?

The transmitter uses a 10/30 Vdc power supply.

How is the conductivity scale selected?

The scale is selected with dip-switches S2 and S3 according to the manual's table.

How do I enable automatic temperature compensation?

Remove the jumpers from terminals 3-4 and 5-6, then connect the RTD Pt100 between terminals 4-5-6.

What type of output does the transmitter provide?

It provides an isolated 4/20 mA output proportional to conductivity.