This document describes the user manual for the nitrogen generator model N2-Bora from MATHESON TRI-GAS 166 Keystone Drive Montgomeryville, PA 18936.
This Matheson Tri-Gas user manual covers the N2-Bora nitrogen generator, which produces pure nitrogen by filtering compressed air through self-regenerating carbon molecular sieve tubes.
It provides specifications, installation steps, initial start-up and operating instructions, routine maintenance procedures, spare parts information, and FCC compliance notes for the generator.
| Nitrogen flow rate | 0-500, 0-750, or 0-1300 cc/min at STP |
|---|---|
| Max outlet pressure | 5 bar (73 psi) |
| Purity | 99.999%+ (O2 < 10 ppm) |
| Weight | 18 kg |
| Input voltage | 120 or 240V / 50 or 60Hz |
| Fuse | 4A or 2A |
| Pressure accuracy | 0.1 bar (± 0.5 %) |
| Display | Graphic display, 128 x 64 pixels |
| Index of protection | IP2x |
| Operating temperature | 15°C to +40°C |
The outlet pressure is adjusted with a forward pressure regulator and is by default set to 5 bar.
The carbon molecular sieve (CMS) must be replaced every 24000 hours.
If the password is forgotten, the default password 04122 can be submitted.
Purity is 99.999%+ with O2 below 10 ppm.