The Mettler-Toledo 8142-2009 is a non-automatic weighing instrument that complies with CE certification standards.
The Mettler-Toledo 8142-2009 is an industrial digital indicator intended for use with analog load cells only. The 8142 is available in single display, dual display, and dual display-advanced versions.
This technical manual covers installation instructions, programming and calibration, operating instructions, hardware setup, maintenance, spare parts and accessories, interconnect drawings, and a parts catalog. It is written as a guide for individuals who have received technical training in servicing the product.
| Display | 6 digit, seven segment, blue-green, fluorescent display |
|---|---|
| Load cell input | 2 mV/V or 3 mV/V, jumper selectable |
| Excitation voltage | +12.5 VDC or +5 VDC |
| Displayed increments | 600 to 50,000 increments |
| Increment size | 0.00001 to 100 |
| Motion detection | ±0.5 to ±3 increments |
| Pushbutton zero range | ±2% or ±20% of programmed scale capacity |
| Units | pounds (lb) or kilograms (kg) |
| Auto zero maintenance | ±0.5 increments to ±3 increments |
| Span range for 2 mV/V load cells | 3 mV to 30 mV |
| Initial dead load offset for 2 mV/V load cells | up to 27 mV |
The 8142 is intended for use with analog load cells only.
The 8142 is available in single display, dual display, and dual display-advanced versions.
The total number of displayed increments is selectable from 600 to 50,000.
It supplies +12.5 VDC or +5 VDC excitation voltage.