Additel ADT875 Series Dry Well Calibrator User Manual

Additel ADT875 Series Dry Well Calibrator User Manual


Additel ADT875 Series Dry Well Calibrator User Manual - cover page
Brand
Additel
Category
Calibration Kit
Document type
User Manual
Language
English
Pages
182
File format
PDF
File size
4 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
2F148BD8FF8E090B69FF6BCB826384AF

About this manual

This manual covers the Additel ADT875 Series dry well calibrator, including model information, basic structure, features, environmental conditions, technical specifications, and standard equipment. It also explains operation, system setup, task management, applications, and the ADT875 SCPI command list.

The documented models include ADT875PC, ADT875-155, ADT875-350, and ADT875-660, with dry well temperature ranges from -40 C to 155 C, 33 C to 350 C, and 33 C to 660 C. Electrical measurement functions are documented for ADT875PC, and transmitter measurement includes HART transmitter measurement.

What's inside

  • 1. Introduction
  • 2. Operation
  • 3. System Setup
  • 4. Task
  • 5. Application
  • Appendix 1: ADT875 SCPI Command List

Specifications

Display accuracy±0.18 C at Full Range; ±0.2 C at Full Range; ±0.3 C at 33 C; ±0.3 C at 420 C; ±0.5 C at 660 C
Stability (30 min)±0.01 C at Full Range; ±0.02 C at Full Range; ±0.02 C at 33 C; ±0.03 C at 50 C; ±0.04 C at 420 C; ±0.04 C at 660 C
Axial uniformity at 60 mm (2.4 in)±0.07 C at Full Range; ±0.04 C at 33 C; ±0.05 C at 33 C; ±0.1 C at 200 C; ±0.3 C at 420 C; ±0.2 C at 350 C; ±0.5 C at 660 C
Radial uniformity±0.01 C at Full Range; ±0.01 C at 33 C; ±0.02 C at 33 C; ±0.015 C at 200 C; ±0.05 C at 420 C; ±0.02 C at 350 C; ±0.1 C at 660 C
Loading effect±0.1 C (Display Sensor); ±0.15 C (Display Sensor); ±0.15 C (Display Sensor); ±0.02 C (External Sensor); ±0.015 C (External Sensor); ±0.025 C (External Sensor)
Hysteresis (Display Sensor)0.025 C; 0.03 C; 0.1 C
Immersion depth150 mm (5.9 in)
Readout accuracy for 100 ohm PRT±0.009 C at -40 C to ±0.042 C at 660 C
RTD measurement accuracy (excl sensor)0 Ω to 25 Ω: 0.002 Ω; 25 Ω to 400 Ω: 0.008% RD; 400 Ω to 4K Ω: 0.004% RD
Reference resistance range0 Ω to 400 Ω
Reference resistance accuracy0 Ω to 50 Ω: 0.002 Ω; 50 Ω to 400 Ω: 0.004% RD
RTD channels2
Reference probe connection6-pin lemo smart connector
Reference characterizationsITS-90, CVD, IEC-751
Readout resolution1 mΩ

Frequently asked questions

Which models are included in the ADT875 Series?

The dry well specifications table lists ADT875PC, ADT875-155, ADT875-350, and ADT875-660.

What is the immersion depth of the dry well?

The immersion depth is 150 mm (5.9 in).

Which model supports electrical measurement?

Electrical measurement functions are only for ADT875PC.

Does the calibrator support HART transmitter measurement?

Transmitter measurement includes HART transmitter measurement.

Download PDF (4 MB)