Additel ADT875 Series Dry Well Calibrator User Manual
This manual covers the Additel ADT875 Series dry well calibrator, including model information, basic structure, features, environmental conditions, technical specifications, and standard equipment. It also explains operation, system setup, task management, applications, and the ADT875 SCPI command list.
The documented models include ADT875PC, ADT875-155, ADT875-350, and ADT875-660, with dry well temperature ranges from -40 C to 155 C, 33 C to 350 C, and 33 C to 660 C. Electrical measurement functions are documented for ADT875PC, and transmitter measurement includes HART transmitter measurement.
| Display accuracy | ±0.18 C at Full Range; ±0.2 C at Full Range; ±0.3 C at 33 C; ±0.3 C at 420 C; ±0.5 C at 660 C |
|---|---|
| Stability (30 min) | ±0.01 C at Full Range; ±0.02 C at Full Range; ±0.02 C at 33 C; ±0.03 C at 50 C; ±0.04 C at 420 C; ±0.04 C at 660 C |
| Axial uniformity at 60 mm (2.4 in) | ±0.07 C at Full Range; ±0.04 C at 33 C; ±0.05 C at 33 C; ±0.1 C at 200 C; ±0.3 C at 420 C; ±0.2 C at 350 C; ±0.5 C at 660 C |
| Radial uniformity | ±0.01 C at Full Range; ±0.01 C at 33 C; ±0.02 C at 33 C; ±0.015 C at 200 C; ±0.05 C at 420 C; ±0.02 C at 350 C; ±0.1 C at 660 C |
| Loading effect | ±0.1 C (Display Sensor); ±0.15 C (Display Sensor); ±0.15 C (Display Sensor); ±0.02 C (External Sensor); ±0.015 C (External Sensor); ±0.025 C (External Sensor) |
| Hysteresis (Display Sensor) | 0.025 C; 0.03 C; 0.1 C |
| Immersion depth | 150 mm (5.9 in) |
| Readout accuracy for 100 ohm PRT | ±0.009 C at -40 C to ±0.042 C at 660 C |
| RTD measurement accuracy (excl sensor) | 0 Ω to 25 Ω: 0.002 Ω; 25 Ω to 400 Ω: 0.008% RD; 400 Ω to 4K Ω: 0.004% RD |
| Reference resistance range | 0 Ω to 400 Ω |
| Reference resistance accuracy | 0 Ω to 50 Ω: 0.002 Ω; 50 Ω to 400 Ω: 0.004% RD |
| RTD channels | 2 |
| Reference probe connection | 6-pin lemo smart connector |
| Reference characterizations | ITS-90, CVD, IEC-751 |
| Readout resolution | 1 mΩ |
The dry well specifications table lists ADT875PC, ADT875-155, ADT875-350, and ADT875-660.
The immersion depth is 150 mm (5.9 in).
Electrical measurement functions are only for ADT875PC.
Transmitter measurement includes HART transmitter measurement.