KSA264Ge11 96T x 5 ELISA Kit DIY Materials For Pentosidine (PTD) Organism Species: Pan-species (General)
This instruction manual covers the Cloud-Clone KSA264Ge11 96T x 5 ELISA Kit DIY Materials for Pentosidine (PTD), intended for pan-species (general) use. The kit provides materials for developing a competitive inhibition ELISA to measure PTD in serum, plasma and other biological fluids in vitro, and contains sufficient materials for preparation of at least five 96-well plates. It is for in vitro and research use only, not for use in clinical diagnostic procedures.
The manual lists the reagents provided (Capture Antibody, Biontin-labeled Competitor, Standard, Streptavidin-HRP, TMB Substrate and five 96-well plates), storage conditions, reagent preparation with dilution instructions, a step-by-step assay protocol including plate preparation and measurement at 450 nm, plus reference detection range, specificity notes and a typical standard curve.
| Kit format | 96T x 5 (at least five 96-well plates) |
|---|---|
| Detection range (reference) | 6.17-500 ng/mL |
| Standard concentration | 500 ng/mL |
| Measurement wavelength | 450 nm |
| Shelf life | 12 months |
| Storage | Antibodies, Standard, Biontin-labeled Competitor and Streptavidin-HRP at -20°C; TMB at 4°C; plates at room temperature |
| Required support pack | Assay Kit DIY Support Pack 2 (Catalog: IS050) |
The detection range in the manufacturer's lab is 6.17-500 ng/mL.
Antibodies, Standard, Biontin-labeled Competitor and Streptavidin-HRP are stored at -20°C, TMB at 4°C, and the 96-well plates at room temperature. Reagents are valid for 12 months and stable for one month after opening at 4°C.
It is for the development of a competitive inhibition ELISA to measure PTD in serum, plasma and other biological fluids in vitro, for research use only and not for clinical diagnostic procedures.
No significant cross-reactivity or interference between PTD and analogues was observed.