This quick start guide is written to help you easily confirm the operation method and display contents.
This quick start manual from ULVAC is written to let users quickly confirm the operating methods and display contents of the GI-M2 and GI-D7 metal ionization vacuum gauges. It applies to GI-M2 and GI-D7 units from the serial numbers stated in the document onward. The full instruction manual with detailed usage, precautions and safety information can be downloaded from the ULVAC website.
The document explains setting items such as range switching, sensitivity coefficient, setpoints 1 and 2, DEGAS operation (manual and auto modes with a settable time of 01-99 minutes), filament 1/2 selection and analog output modes, including linear, log and GI-D6/GI-TL3 compatible outputs with their pressure-to-voltage formulas. It also describes the names and functions of the front-panel switches and indicators, external input/output signals, RS-232C communication, basic specifications, standard accessories and warranty terms. This manual is written in Chinese.
| Power supply | AC100V ±10V, 50/60 Hz |
|---|---|
| Sampling time | 100 ms (display and output) |
| Contact capacity (setpoint output) | AC125V 0.5A max / DC24V 1.0A max |
| Operating humidity range | 15 to 80% (no condensation) |
| Dimensions | 240 mm (W) × 380 mm (D) × 99 mm (H) |
| DEGAS auto OFF settable time | 01 to 99 minutes |
The full manual can be downloaded from the ULVAC showcase website listed in the manual.
It applies to GI-M2 and GI-D7 ionization vacuum gauges from the serial numbers stated in the document onward.
In AUTO mode the DEGAS time can be set from 01 to 99 minutes, after which DEGAS turns off automatically.
The warranty period is one year from the date of purchase, covering the sensor unit of the instrument.