UM-1 and UM-2 series ultrasonic thickness gauges adopt ultrasonic measurement principle and are ultra-small measuring instruments. It can measure quickly, nondestructively and accurately. This instrument can be widely used in manufacturing, metal processing, chemical industry, commercial inspection and other testing fields, and is suitable for measuring the thickness of various materials that enable ultrasonic waves to propagate at a constant speed inside them and reflect from their backsides.
Yushi Xianfeng UM-1, UM-1D, UM-2 and UM-2D ultrasonic thickness gauges measure thickness by sending ultrasonic pulses through a couplant and timing their return from material interfaces. The manual covers instrument overview, technical parameters, main functions, measurement steps, storage and review of thickness values, deletion, system and function settings, measurement application techniques, and maintenance.
Supported functions include coating penetration on D models, quick and two-point calibration, storage of 500 values in five files, thickness alarms, minimum value capture, millimeter and inch units, five stored sound velocities, backlight, low-voltage indication and automatic shutoff. The document also lists material sound velocities and probe measurement ranges. This manual is written in Chinese.
| Display | 128x64 dot-matrix LCD |
|---|---|
| Standard mode measurement range | 0.8 mm to 300 mm |
| Coating mode measurement range | 3 mm to 18 mm |
| Indication error (UM-1/UM-1D) | ±0.1 mm below 10 mm; ±(1%H+0.1) mm from 10 mm to upper limit |
| Indication error (UM-2/UM-2D) | ±0.05 mm below 10 mm; ±(0.5%H+0.01) mm from 10 mm to upper limit |
| Repeatability (UM-1/UM-1D) | ±0.1 mm |
| Repeatability (UM-2/UM-2D) | ±0.05 mm |
| Display resolution (UM-1/UM-1D) | 0.1 mm |
| Display resolution (UM-2/UM-2D) | 0.01/0.1 mm selectable below 100.0 mm; 0.1 mm at or above 100.0 mm |
| Measurement frequency | 4 Hz |
| Sound velocity adjustment range | 1000 to 9999 m/s |
| Operating environment | -10 C to 60 C |
| Power | Two AA alkaline batteries |
| Dimensions | 149 x 73 x 32 mm |
| Weight | 160 g without battery |
| Thickness value storage | 500 values in five files |
| Stored sound velocities | 5 |
| Automatic shutoff | After 5 minutes of inactivity |
It can store 500 thickness values, divided into five files of 100 values each.
The standard mode measurement range is 0.8 mm to 300 mm, depending on the probe and material.
Quick calibration should be performed after changing the probe and can be done before each measurement if necessary.
In the stored-data review state, press ENTER to delete the currently displayed value.