Yushi Xianfeng UM-1 and UM-2 Series Ultrasonic Thickness Gauge User Manual

UM-1 and UM-2 series ultrasonic thickness gauges adopt ultrasonic measurement principle and are ultra-small measuring instruments. It can measure quickly, nondestructively and accurately. This instrument can be widely used in manufacturing, metal processing, chemical industry, commercial inspection and other testing fields, and is suitable for measuring the thickness of various materials that enable ultrasonic waves to propagate at a constant speed inside them and reflect from their backsides.


Yushi Xianfeng UM-1 and UM-2 Series Ultrasonic Thickness Gauge User Manual - cover page
Brand
Yushi Xianfeng
Category
Other device
Document type
User Manual
Language
Chinese
Pages
14
File format
PDF
File size
364 KB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
D860593D0B97F98D57F1A1F6195DE543

About this manual

Yushi Xianfeng UM-1, UM-1D, UM-2 and UM-2D ultrasonic thickness gauges measure thickness by sending ultrasonic pulses through a couplant and timing their return from material interfaces. The manual covers instrument overview, technical parameters, main functions, measurement steps, storage and review of thickness values, deletion, system and function settings, measurement application techniques, and maintenance.

Supported functions include coating penetration on D models, quick and two-point calibration, storage of 500 values in five files, thickness alarms, minimum value capture, millimeter and inch units, five stored sound velocities, backlight, low-voltage indication and automatic shutoff. The document also lists material sound velocities and probe measurement ranges. This manual is written in Chinese.

What's inside

  • Instrument overview
  • Measurement principle
  • Coating penetration function
  • Basic configuration and part names
  • Technical parameters
  • Main functions
  • Measurement steps
  • Thickness value storage and review
  • Deletion operations
  • System and function settings

Specifications

Display128x64 dot-matrix LCD
Standard mode measurement range0.8 mm to 300 mm
Coating mode measurement range3 mm to 18 mm
Indication error (UM-1/UM-1D)±0.1 mm below 10 mm; ±(1%H+0.1) mm from 10 mm to upper limit
Indication error (UM-2/UM-2D)±0.05 mm below 10 mm; ±(0.5%H+0.01) mm from 10 mm to upper limit
Repeatability (UM-1/UM-1D)±0.1 mm
Repeatability (UM-2/UM-2D)±0.05 mm
Display resolution (UM-1/UM-1D)0.1 mm
Display resolution (UM-2/UM-2D)0.01/0.1 mm selectable below 100.0 mm; 0.1 mm at or above 100.0 mm
Measurement frequency4 Hz
Sound velocity adjustment range1000 to 9999 m/s
Operating environment-10 C to 60 C
PowerTwo AA alkaline batteries
Dimensions149 x 73 x 32 mm
Weight160 g without battery
Thickness value storage500 values in five files
Stored sound velocities5
Automatic shutoffAfter 5 minutes of inactivity

Frequently asked questions

How many thickness values can the gauge store?

It can store 500 thickness values, divided into five files of 100 values each.

What is the standard mode measurement range?

The standard mode measurement range is 0.8 mm to 300 mm, depending on the probe and material.

When should quick calibration be performed?

Quick calibration should be performed after changing the probe and can be done before each measurement if necessary.

How can a single stored thickness value be deleted?

In the stored-data review state, press ENTER to delete the currently displayed value.