Fuli GC9790 Gas Chromatograph Installation and Operation Manual

The Fuli GC9790 Gas Chromatograph is a versatile and high-performance single detector series instrument.


Fuli GC9790 Gas Chromatograph Installation and Operation Manual - cover page
Brand
Fuli
Category
Analyzer
Document type
User Manual
Language
Chinese
Pages
53
File format
PDF
File size
1.1 MB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
6D48606FA75EE114DDBE3790939D05B4

About this manual

This manual covers the Fuli GC9790 gas chromatograph, a GC97 series general-purpose, multi-purpose, high-performance single-detector instrument made by Zhejiang Wenling Fuli Analytical Instrument Co., Ltd. (FULI).

It includes power and gas source requirements, FID and TCD operating rules, technical specifications, installation steps, operation instructions, gas path system, injector, detector, maintenance, and common fault troubleshooting. This manual is written in Chinese.

What's inside

  • Overview
  • Instrument technical specifications
  • Instrument completeness
  • Instrument working principle
  • Installation
  • Operation
  • Gas path
  • Injector
  • FID hydrogen flame ionization detector
  • Instrument maintenance and common fault handling

Specifications

Dimensions490×460×480 mm (L×H×W)
Weight40 kg
Column oven temperature rangeAmbient +10°C to 350°C
FID detection limitNot greater than 2×10^-12 g/s
Supply voltage220±10%
Ambient temperature5°C to 35°C

Frequently asked questions

What are the power requirements?

The instrument requires 220V±10% power, 50±0.5Hz frequency, grounding resistance below 0.1Ω, and power greater than 2000 W.

What is the FID detection limit?

The FID detection limit is not greater than 2×10^-12 g/s [n-C16].

How much does the instrument weigh?

The instrument weighs 40 kg.