Iris Diagnostics iRICELL Urine Analyzer: Automated vs Manual Urinalysis Comparison Study

The study showed that the automated urine analyzer was superior to the manual method in both accuracy and efficiency.


Iris Diagnostics iRICELL Urine Analyzer: Automated vs Manual Urinalysis Comparison Study - cover page
Brand
Iris Diagnostics
Category
Analyzer
Document type
Other
Language
Spanish
Pages
10
File format
PDF
File size
371 KB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
F2B5C3BF9C21608BDD83B3FB7B14D979

About this manual

This document is a scientific study by laboratory staff at Sanatorio Allende in Córdoba, Argentina, comparing the automated iRICELL urine analysis system from Iris Diagnostics with the conventional manual urinalysis technique. The iRICELL system consists of two modules: the iChem VELOCITY for chemistry and the iQ200 ELITE for microscopy.

The authors analyzed 200 urine samples from outpatients, inpatients and emergency patients, processing each by both methods simultaneously. They report acceptable concordance and correlation between the two methods for the parameters studied, including cells, red blood cells, leukocytes, pH, nitrites, hemoglobin, color, turbidity, specific gravity, proteinuria and glucosuria, and conclude that automation improved the quality of the urinalysis in their laboratory. This manual is written in Spanish.

What's inside

  • Abstract
  • Introduction
  • Materials and Methods
  • Results
  • Discussion
  • Conclusion

Frequently asked questions

Which modules make up the iRICELL urine analysis system?

The iRICELL system consists of two modules: iChem VELOCITY for chemistry and iQ200 ELITE for microscopy.

How many urine samples were analyzed in the comparison?

200 urine samples from ambulatory, hospitalized and emergency patients were processed simultaneously by both methods.

What concordance was found between the automated and manual methods?

Kappa was above 0.7 for cells and red blood cells, above 0.8 for pH, nitrites, pathological sediment and leukocytes, and above 0.9 for color, turbidity and hemoglobin.

Which reagent strips and microscope were used for the manual technique?

The manual method used SIEMENS Multistix 10 SG reagent strips and a ZEISS Primo Star microscope.

Download PDF (371 KB)