The HYG2670P-I withstand voltage tester is a high-performance withstand voltage tester. It can set the output voltage, test alarm value, test time, etc. by software, and has a variety of automatic test functions. This series of testers has high accuracy of high voltage output (better than 2%) and high accuracy of leakage current test (better than 2%). It can be configured with RS-232 computer interface and remote control interface, which is suitable for both production line rapid流水testing and laboratory multi-function high-precision requirements.
The HYG2670P-I is a programmable withstand voltage tester from Wuhan Huanneng Yangguang Electric. The manual begins with safety rules and installation points: the instrument runs from a single-phase 220 V/50 Hz supply with a 2 A slow-blow fuse, must be grounded through its three-core power cord, and its safety requirements are stated to follow GB4706.1 and IEC1010-1. Output voltage, alarm values and test times can be set in software.
The overview describes the front panel (power switch, setting, decrease, increase and exit keys, high- and low-voltage output terminals, start and stop buttons, and a 16×2 character LCD) and the rear panel (ground terminal, optional RS-232 or RS-485 serial port, 220 VAC input with fuse and spare fuse, remote switch, nameplate and ventilation holes). The manual also lists standard accessories, general specifications, test precautions, daily maintenance and troubleshooting, a 12-month quality warranty, and an appendix on withstand voltage calibration. This manual is written in Chinese.
| Input voltage | 220 Vac ±10% |
|---|---|
| Input frequency | 50/60 Hz |
| Fuse | 2 A / 250 V slow-blow |
| Display | 16×2 blue character LCD |
| Communication interface | RS232 (optional) |
| High-voltage output accuracy | Better than 2% |
| Leakage current measurement accuracy | Better than 2% |
| Operating environment | 0–40 °C, 20–80% RH |
| Dimensions | 280 mm (W) × 89 mm (H) × 380 mm (D) |
| Calibration method | Software calibration via front-panel parameter entry |
It uses a single-phase 220 V/50 Hz supply, with the input rated at 220 Vac ±10% and 50/60 Hz.
The manual states that the measurement principle meets the safety performance requirements of GB4706.1, GB3883.1 and GB4943.
The accessory list includes the user manual, certificate and warranty card, high-voltage test probe 8301, return line 8311, remote interface 8002, 220 V input power cord 8122 and a 2 A slow-blow fuse.
The troubleshooting table gives a blown or badly contacting fuse in the power socket; replace it with a fuse of the same rating.