APC-4013 Laser Dust Particle Counter is a professional measuring instrument for detecting the cleanliness level of air in clean rooms and workshops. It complies with the technical requirements of the JJF1190-2008 Dust Particle Counter Calibration Specification issued by the State Bureau of Metrology. The functions of the whole machine are controlled and processed by a microcomputer, and the detection results can be printed directly.
The APC-4013 is a laser dust particle counter from Shanghai Ruihong Detection Technology (Ruihong) with a large LCD screen. The leaflet describes it as a professional measuring instrument for checking the air cleanliness class of cleanroom workshops, built to the technical requirements of calibration specification JJF1190-2008, controlled by a microcomputer and able to print measurement results directly.
A single sample measures six particle size channels at once — 0.3, 0.5, 1.0, 2.0, 3.0 and 5.0 μm — with a sampling volume of 2.83 L/min (0.1 cfm) and UCL calculation at 95% confidence. Listed uses include pharmaceutical plants, drug inspection institutes and hospital operating rooms, semiconductor factories and precision machining, food hygiene work, filter manufacturers checking filter quality and efficiency, and cleanroom testing in optics and aerospace. The page lists the main specifications and the accessories supplied with the unit. This manual is written in Chinese.
| Particle size channels | Six channels: 0.3, 0.5, 1.0, 2.0, 3.0, 5.0 μm |
|---|---|
| Sampling volume | 2.83 L/min (0.1 cfm) |
| Detection range | Class 100 to Class 300,000 |
| Test period | 1 min to 10 min (10 settings) |
| Self-cleaning time | ≤10 min |
| Operating environment | Temperature 10–35 °C, relative humidity ≤75% |
| Vacuum source | Built-in air pump with flow control |
| Light source and life | Semiconductor laser, >30000 h |
| Printing function | Built-in printer; prints counts in ALL (actual value) or 209E mode |
| Dimensions (W×L×H) | 240×300×128 mm |
| Screen size (W×H) | 110×75 mm |
| Weight | 3.6 kg |
| Power supply | 220 V ±10%, 50 Hz ±2 Hz |
It has six size channels — 0.3, 0.5, 1.0, 2.0, 3.0 and 5.0 μm — and a single sample measures all six at the same time.
The sampling volume is 2.83 L/min (0.1 cfm).
The accessories include an aluminum alloy case, operation manual, factory test report, power cord, sampling stand, printer paper and fuse.
It is specified for temperatures of 10 to 35 °C and a relative humidity of 75% or less.