This product is a plant sucrose content detection kit with the model number BC2465 and the specification of 100T/96S.
This two-page instruction manual covers the Solarbio BC2465 plant sucrose content detection kit (micro method, specification 100T/96S). It lists the kit components and their storage conditions, gives reagent preparation instructions, and explains the assay principle: reducing sugars are destroyed by heating with alkali, sucrose is then hydrolyzed to glucose and fructose, and fructose reacts with resorcinol to form a colored substance with a characteristic absorption peak at 480 nm.
The manual describes sample processing (weighing 0.1 g of sample, extraction, an 80 °C water bath, centrifugation, and decolorization), the measurement procedure using blank, standard, and assay tubes read at 480 nm, and formulas for calculating sucrose content per protein concentration or per sample mass. It also states the minimum detection limit (0.0316 mg/mL), the linear range (0.039-12 mg/mL), the instruments to be provided by the user, and related Solarbio sucrose metabolism kits. This manual is written in Chinese.
| Specification | 100T/96S |
|---|---|
| Minimum Detection Limit | 0.0316 mg/mL |
| Linear Range | 0.039-12 mg/mL |
| Measurement Wavelength | 480 nm |
| Standard Tube Concentration | 1 mg/mL |
| Extraction Solution | Liquid 100 mL × 1 bottle, stored at 4 °C |
| Sample Amount | 0.1 g |
Reducing sugars are destroyed by heating the sample with alkali, sucrose is hydrolyzed to glucose and fructose, and the fructose reacts with resorcinol to form a colored substance read at 480 nm.
The linear range is 0.039-12 mg/mL, with a minimum detection limit of 0.0316 mg/mL.
A visible spectrophotometer or microplate reader, water bath, adjustable pipette, micro glass cuvette or 96-well plate, mortar or homogenizer, and distilled water.
Two formulas are given: per protein concentration, (A3-A1)/(A2-A1)/Cpr, and per sample mass, (A3-A1)/(A2-A1)/W.