GHC-9600 Transformer Oil Gas Chromatograph adopts the three-detector process recommended by national standards, and can complete the full analysis of the content of dissolved gas components (including hydrogen, oxygen, methane, ethylene, ethane, acetylene, carbon monoxide and carbon dioxide) in the insulating oil with one injection.
The GHC-9600 is an oil chromatograph analyzer for power systems produced by Wuhan Guoce Huaneng Electric Co., Ltd. (武汉国测华能电气有限公司). Following the national-standard three-detector flow, a single injection completes the full analysis of gases dissolved in insulating oil — hydrogen, oxygen, methane, ethylene, ethane, acetylene, carbon monoxide and carbon dioxide — which generation and supply utilities use to detect hidden overheating or discharge faults in oil-filled equipment. The system also determines the gas content of insulating oil, with a flow design that complies with the power industry standard DL/T 703-1999.
A high-sensitivity thermal conductivity detector, a hydrogen flame ionization detector and a nickel catalyst conversion furnace enable measurement of nine gas components, including nitrogen. The manual lists the instrument configuration (chromatograph mainframe with FID and TCD detectors, conversion furnace, dedicated workstation, shaker, gas sources, standard gases and accessories), the technical parameters, and features such as networked data transmission with remote control and diagnosis, a 5.7-inch LCD switchable between Chinese and English, carrier-gas failure protection for the TCD, 0.1°C four-channel temperature control, self-diagnosis and anti-pollution chromatographic columns. This manual is written in Chinese.
| Sample volume | 1 mL per injection |
|---|---|
| Minimum detection concentration | H2 2, O2 10, N2 20, CO 1, CO2 5, CH4 0.1, C2H4 0.1, C2H6 0.1, C2H2 0.1 μL/L |
| Detectors | TCD and FID with nickel catalyst conversion furnace |
| Ambient temperature | 0°C to 40°C |
| Temperature control range | Room temperature +5°C to 400°C |
| Temperature control accuracy | ±0.1°C |
| Stabilization time | Under 30 min |
| Power | 1.9 kW |
| Dimensions | 660 × 550 × 585 mm |
| Display | 5.7-inch LCD, 13 lines, Chinese/English switchable |
It measures nine dissolved gas components: hydrogen, oxygen, nitrogen, carbon monoxide, carbon dioxide, methane, ethane, ethylene and acetylene.
It uses a high-sensitivity thermal conductivity detector (TCD) and a hydrogen flame ionization detector (FID), plus a nickel catalyst conversion furnace.
Its flow design complies with the Chinese power industry standard DL/T 703-1999 for gas chromatographic determination of gas content in insulating oil.
One injection of 1 mL completes the full analysis of the dissolved gas components.