The MA series moisture meters produced by Nanjing Bonita Scientific Instrument Co., Ltd. have the following models: MA50-2H, MA100-2H, MA50-3H, MA100-3H, and MA110-3H. The main technical parameters include range, accuracy level, weighing range, actual scale value, verification scale value, moisture determination accuracy, moisture content determination readability, temperature control tolerance, heating temperature range setting, heating time range setting, analysis method, display parameters, moisture content determination range, product size, outer box size, net weight/gross weight, etc.
This is the operating manual for the MA series moisture analyzers made by Nanjing Bonita Scientific Instrument Co., Ltd. It covers the working environment and safety precautions, main technical parameters, packing list, installation and adjustment, and step-by-step operating instructions.
Later sections describe the data interface (serial port settings), maintenance and care, and a troubleshooting table with fault symptoms, causes and solutions. The manual also explains weighing calibration (single-point and multi-point), drying temperature settings, the three ending modes (manual, timed, automatic), the nine displayed moisture parameters, record storage and printing via the RS232 interface. This manual is written in Chinese.
| Maximum capacity | 50g / 100g / 50g / 100g / 110g (MA50-2H / MA100-2H / MA50-3H / MA100-3H / MA110-3H) |
|---|---|
| Actual scale interval | 0.01g / 1mg |
| Verification scale interval | 0.1g / 0.01g |
| Moisture measurement accuracy | ±0.5% (sample ≥ 2g) |
| Moisture content readability | 0.02%~0.1% (sample ≥ 2g) |
| Temperature control tolerance | ±1℃ |
| Heating temperature range | 60~200℃ (adjustable in 1℃ steps) |
| Heating time range | 0–99 minutes (adjustable in 1-minute steps) |
| Analysis methods | Manual, timed, automatic |
| Moisture content measurement range | 0%~100% |
| Product dimensions | 370mm×215mm×195mm |
| Net weight / gross weight | 3.5kg / 4.5kg |
| Power supply | 220V, 50Hz±1Hz, approx. 300W |
| Working environment | 15℃~30℃, temperature fluctuation < 5℃/h, relative humidity 50%~75% |
| Serial interface | Baud rate 9600, no parity |
The manual recommends warming up for 20-30 minutes before performing drying operations.
Typical samples should be 5g to 10g, spread evenly on the pan and tared before weighing; powder samples give the best accuracy.
Press the calibration key, select weight calibration and then range calibration, and place the 20g, 50g or 100g calibration weight matching the displayed value.
Up to ten sets of dried sample results can be saved; when full, the newest result replaces the first record.