This document provides a selection guide for ATEX (GasEX and DustEX) filters.
This one-page ENOTEC selection guide covers filter choice for ATEX (GasEX and DustEX) systems. It states that neither the filter material nor the filter head construction can be chosen to be ATEX compliant, so ATEX systems are not configurable in this respect.
The guide then describes filter head configuration for horizontal and vertical probe installation and compares the filter materials ENOTEC offers: sintered metal (1.4404), ceramic (Al2O3), basalt and alloy (2.4816), with corrosion behavior, dust load, mechanical strength, temperature and application examples. An additional sintered metal ring for faster gas exchange is noted as available only for KES-1000 (OXITEC Economy) and KES-500X probes.
| ATEX filter head/material configurability | Not configurable for ATEX compliance |
|---|---|
| Recommended horizontal probe mounting angle | -3° to -5° |
| Vertical probe mounting angle | -40° to -90° |
| Filter materials offered | Sintered metal, ceramic, basalt, alloy |
| Sintered metal material | 1.4404 |
| Ceramic material | Al2O3 |
| Alloy material | 2.4816 |
No. Neither the filter material nor the filter head construction can be chosen to be ATEX compliant, so ATEX systems are not configurable in this aspect.
The filter head of vertical installation should be used, so that a fast response time to process gas changes is provided.
A mounting angle of -3° to -5° is recommended; the standard filter head is suitable for this installation.