ENOTEC GasEX and DustEX ATEX Filter Selection Guide

This document provides a selection guide for ATEX (GasEX and DustEX) filters.


ENOTEC GasEX and DustEX ATEX Filter Selection Guide - cover page
Brand
Enotec
Category
Analyzers
Document type
Product Leaflet
Language
English
Pages
1
File format
PDF
File size
408 KB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
BBE78FD3AED8210A93D0E58DB7132D96

About this manual

This one-page ENOTEC selection guide covers filter choice for ATEX (GasEX and DustEX) systems. It states that neither the filter material nor the filter head construction can be chosen to be ATEX compliant, so ATEX systems are not configurable in this respect.

The guide then describes filter head configuration for horizontal and vertical probe installation and compares the filter materials ENOTEC offers: sintered metal (1.4404), ceramic (Al2O3), basalt and alloy (2.4816), with corrosion behavior, dust load, mechanical strength, temperature and application examples. An additional sintered metal ring for faster gas exchange is noted as available only for KES-1000 (OXITEC Economy) and KES-500X probes.

What's inside

  • Filter head configuration
  • Horizontal installation of the probe
  • Vertical installation of the probe
  • Filter material
  • Additional sintered metal ring for faster gas exchange

Specifications

ATEX filter head/material configurabilityNot configurable for ATEX compliance
Recommended horizontal probe mounting angle-3° to -5°
Vertical probe mounting angle-40° to -90°
Filter materials offeredSintered metal, ceramic, basalt, alloy
Sintered metal material1.4404
Ceramic materialAl2O3
Alloy material2.4816

Frequently asked questions

Can the filter material or filter head be chosen for ATEX compliance?

No. Neither the filter material nor the filter head construction can be chosen to be ATEX compliant, so ATEX systems are not configurable in this aspect.

Which filter head should be used for vertical probe installation?

The filter head of vertical installation should be used, so that a fast response time to process gas changes is provided.

What mounting angle is recommended for horizontal probe installation?

A mounting angle of -3° to -5° is recommended; the standard filter head is suitable for this installation.