The GS-2 is a gas sampling glass tube used to collect the headspace gas of a container (such as a can) by opening it in water. This method is suitable for measuring samples with low gas content, such as when sufficient gas cannot be collected with a single sample.
The Iijima GS-2 Gas Sampling Glass Tube is designed to collect headspace gas from containers, such as cans, by opening them underwater. This method is specifically suitable for measuring samples with low amounts of gas where a single sample might not be sufficient. The manual details the necessary precautions to ensure safe and proper use of the device.
To operate the GS-2, users must prepare specific equipment including a large water tank, a fixing stand, and an oxygen meter (specifically RO-105 series, RO-103S/103, or RO-102). The process involves submerging the glass tube to remove air, attaching a dedicated needle, and then sinking an unopened sample container into the water to capture the gas. The collected gas volume is indicated by a scale on the glass tube.
| Minimum sample gas volume | 6 mL |
|---|---|
| Minimum internal pressure | -30 kPa |
| Recommended gas amount | approx. 10 mL |
Measurements may not be possible if there is less than 6 mL of sample gas.
The GS-2 is compatible with oxygen meter models RO-105 series, RO-103S/103, and RO-102.
The amount of gas in the container is indicated by the scale on the glass tube above the water level.