Hydroinnova CRS-1000 User’s Guide introduces Q-DL-2100 data logger platform, which consists of microcontroller, real-time clock, analog input subsystem, analog output subsystem, digital input/output subsystem, redundant data storage, RS-232 interfaces, SDI-12 interfaces, regulated 5V and 12V outputs, relays, integrated temperature, humidity and pressure sensors, remote communications subsystem.
The Q-DL-2100 is a Hydroinnova data logger and control unit platform documented in Appendix D of the CRS-1000 User's Guide. It consists of a microcontroller with firmware, a real-time clock, analog input and output subsystems, digital I/O, redundant SD and Micro SD storage, RS-232 and SDI-12 interfaces, regulated 5 V and 12 V outputs, relays, integrated temperature/humidity/pressure sensors, and an optional remote communications subsystem with GSM/CDMA or Iridium modems.
Appendix D lists Q-DL-2100 specifications and options, describes the front and back panel layouts, and provides RJ45 connector pinouts. It also gives installation steps and explains configuration through QILogger.INI parameters loaded from the external SD card or internal SD card.
| Program execution rate | 1 min to 1 yr, in 1 minute increments |
|---|---|
| Analog inputs | 4 single-ended inputs; 0 to 5 V range; 2.5 µV resolution (21 bits) |
| Analog outputs | 4 independent 16-bit DAC outputs; 0 to 5 V range; ~0.08 mV resolution |
| Digital I/O ports | 4 ports, each configurable as TTL input, TTL output, or TTL counter |
| Relays | 4 solid-state relays, rated 30 V DC @ 1.2 A each |
| Integrated temperature sensor | Range -55 C to +125 C |
| Barometric pressure sensor | 0.1 mbar resolution |
| SDI-12 sensor support | 2 SDI-12 ports; up to 10 SDI sensors per port |
| Power requirements | 6 to 26 V DC |
| Dimensions | 7 ½ in. x 6 ¾ in. x 2 ¼ in. |
| Real-time clock accuracy | 3 min per year |
| Data storage | One user-removable SD card and one non-removable Micro SD card for redundant data storage |
It has two SDI-12 ports, with up to 10 SDI sensors per port.
It is configured via INI parameters; at power on it attempts to load them from a QILogger.INI text file on the external SD card.
It accepts 6 to 26 V DC.
It adds a GPRS (GSM) integrated cell phone modem.