About this manual

SITRANS Probe LU from Siemens Milltronics is a loop-powered continuous level monitor that uses advanced ultrasonic techniques. The unit combines an electronic component with the transducer and process connection, and its transducer is available in ETFE or PVDF for use with corrosive chemicals. Communication is via HART, and signals are processed using Sonic Intelligence.

This quick start manual covers essential installation, wiring, mounting, and operating information, including RUN and PROGRAM modes, programming, and low-temperature effects. It also lists power, approvals, and mounting requirements; the complete instruction manual contains the full specifications and detailed setup information.

What's inside

  • Specifications
  • Power
  • Approvals
  • Installation
  • Mounting Location Recommendations
  • Mounting Instructions
  • Wiring
  • RUN Mode and PROGRAM Mode
  • Programming
  • Low temperature effects on RUN/PROGRAM modes

Specifications

Process temperature-40 to 85 C (-40 to 185 F)
Ambient temperature surrounding enclosure-40 to 80 C (-40 to 176 F)
Maximum supply voltage30 V DC
Output signal4 to 20 mA
Nominal installation power24 V DC at max. 550 Ohm
Thread types2" NPT, 2" BSP, or PF2/G
Transducer materialsETFE or PVDF
Minimum transducer height above highest level300 mm (1 ft)
Cable jacket strip lengthapproximately 70 mm (2.75 in)
Cover screw torque1.1 to 1.7 N-m (10 to 15 in-lb)

Frequently asked questions

What is SITRANS Probe LU?

It is a loop-powered continuous level monitor that uses advanced ultrasonic techniques.

What transducer materials are available?

The transducer is available in ETFE or PVDF.

What communication protocol does SITRANS Probe LU use?

Communication is via HART.

What is the process temperature range?

The process temperature range is -40 to 85 C (-40 to 185 F).