LINTEC MC-700 Series Mass Flow Controller Instruction Manual

The MC-700 series mass flow controller is a high-precision, high-stability flow controller suitable for flow measurement and control of various gases and liquids. This series of flow controllers uses advanced sensing technology and control algorithms to achieve high-precision, high-stability flow measurement and control.


LINTEC MC-700 Series Mass Flow Controller Instruction Manual - cover page
Brand
Lintec
Category
Mass Flowmeter
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
433 KB
Published
09 November, 2023
Updated
09 November, 2023
MD5 checksum
C5AAFA6E8D123FA3690CB9AEA2E82C98

About this manual

The LINTEC MC-700 Series is a mass flow controller with a variable function (VR) that lets users switch gas type and flow rate using three rotary switches on top of the unit. Models covered are MC-710, MC-720, and MC-730. It uses a LINTEC ambient temperature compensation flow sensor and a digital PID feedback control system that controls valve action so the sensor flow rate output matches the setting value.

The manual explains safety precautions, features, structure, specifications and dimensions, ordering information, analog and digital connections, alarm functions, software switch settings, and operation. It notes that regular maintenance is recommended, with a proofreading frequency of once a year.

What's inside

  • Safety Precautions
  • Summary
  • Features
  • Structure
  • Specification / Dimensions
  • Ordering information
  • Connection
  • Alarm functions
  • Software switch (factory shipped value)
  • Operation

Specifications

Accuracy±1.0%F.S.
Repeatability±0.2%F.S.
Response time1 sec (Typical)
Maximum operating pressure300 kPa(G)
Withstanding pressure1 MPa(G)
Operational temperature/humidity range5 to 50°C / 0 to 80%RH (with no condensation)
Leak integrityLess than 1×10-11Pa·m3/sec (He)
Materials exposed to gasStainless steel 316L, PCTFE, PTFE
Power supply+15VDC±3%: 120mA; -15VDC±3%: 50mA
Digital interfaceRS-485 (RJ45 Jack), 2-wire; maximum connection 32 units
WeightApprox. 1 kg

Frequently asked questions

What power supply is required for the MC-700 Series?

The MC-700 Series requires +15VDC±3% at 120 mA and -15VDC±3% at 50 mA, and ±15VDC must be applied simultaneously.

How are gas type and flow rate changed?

The gas type and flow rate can be changed by using the three rotary switches on the MFC unit.

What digital interface is standard on the MC-700 Series?

RS-485 is standard equipment; an RS-232C digital interface is listed as an option.

How often is proofreading recommended?

Regular maintenance is recommended, and the recommended proofreading frequency is once a year.