The document describes the features and functions of the Pipescan HD family of products.
The Pipescan HD is a hand-pushed magnetic flux leakage (MFL) pipe inspection system using Hall effect sensors. The manual covers three model variants (PS-HD-MFL-127XFLAT, PS-HD-MFL-73X127, and PS-HD-MFL-48X63), compatible acquisition units (Ectane2, Reddy, and Swift-M), and Magnifi, Magnifi GO, or Swift-M GO software.
It describes daily pre-scan checks, setup and deployment, alarm threshold adjustment, scan zone definition, acquisition, marking, report generation, storage and transportation, calibration and reference pipe samples, and maintenance including wheel replacement.
| Technology | Magnetic Flux Leakage / Hall Effect Sensors |
|---|---|
| Method of propulsion | Hand Push |
| Speed | up to 1 m/sec (40 in/sec) |
| OD range - PS-HD-MFL-127XFLAT | 127 mm to flat (5.0 in to flat) |
| OD range - PS-HD-MFL-73X127 | 73 mm to 127 mm (2.875 in to 5 in) |
| OD range - PS-HD-MFL-48X63 | 48 mm to 63 mm (1.9 in to 2.5 in) |
| Max recommended wall thickness - PS-HD-MFL-127XFLAT | Up to 12.7 mm (0.5 in) |
| Max recommended wall thickness - PS-HD-MFL-73X127 | Up to 8.56 mm (0.337 in) |
| Max recommended wall thickness - PS-HD-MFL-48X63 | Up to 6.05 mm (0.238 in) |
| Maximum coating thickness | 6 mm (non magnetic material) |
| Channel spacing | 7 mm (0.27 in) |
| Smallest detectable defect | 2 mm diameter defect at 10% wall loss |
| Maximum surface temperature | 60°C (140°F) |
| Connecting cable | 5 meter (15 feet) standard length |
| Weight - PS-HD-MFL-127XFLAT | 29.5 lbs (13.4 Kg) |
| Weight - PS-HD-MFL-73X127 | 20.5 lbs (9.3 Kg) |
| Weight - PS-HD-MFL-48X63 | 12 lbs (5.5 kg) |
| Transit case | Meets IATA requirements for transporting magnetizable material |
Scanning speed should not exceed 1 m/s (3.28 ft/s).
No, all pipe diameters stated are real outside diameter (OD) measurements, not nominal pipe sizes (NPS).
They should be cross-checked with an alternative method such as UT or a pit gauge.
A keep plate included in the transport case should be used to store the unit.