This document is the instruction manual for the PA-677 70 Amp AC/DC Current Probe.
The TINSE PA-677 is a 70 Amp AC/DC current probe that enables a general-purpose oscilloscope to display AC and DC current signals up to 140 amps peak (50 A RMS). It can also make AC and DC measurements with a multimeter using the recommended MT-246N BNC-to-banana adapter.
This manual covers safety instructions, controls and indicators, basic operation, maintenance, specifications, replaceable parts, and accessories. The probe provides 50 mV/A and 500 mV/A ranges, uses a 9V battery, and has no user-serviceable parts.
| Current ranges | 50/500 mV/A |
|---|---|
| Maximum display current | 140 A peak (50 A RMS) |
| Maximum circuit voltage | 600 VAC |
| DC accuracy, typical | ±3% ±20 mA at 500 mV/A (20 mA to 14 A peak range); ±4% ±200 mA at 50 mV/A (200 mA to 100 A peak range); ±15% max at 50 mV/A (100 A peak to 140 A peak range) |
| Frequency range | DC to 1 MHz (-3 dB) |
| Rise time | 0.35 μS (typical) |
| Battery type and life, typical | 9V NEDA 1604A / IEC 6LR61; 8 hours minimum |
The PA-677 provides 50 mV/A and 500 mV/A ranges, selected with the OFF/Range switch.
It uses one 9V NEDA 1604A (IEC 6LR61) battery with a typical life of 8 hours minimum.
No; the manual warns not to clamp the probe onto circuits with voltages greater than 600 VAC.
Use the recommended MT-246N BNC-to-banana adapter, connect the black lead to COM and the red lead to VΩ, and set the meter to AC or DC volts.