The HSS86.50 PLUS User Manual introduces the product's operation process, carrier and auxiliary gas adjustment, valve switching time setting, method selection and parameter setting, etc.
This user manual for the BCT HSS86.50 PLUS walks through the operating procedure: switch on and let the machine complete its self-test, set the carrier and auxiliary gas pressures, set the valve switching times, select an operating method and its parameters, press START when the instrument shows READY, and switch the power off when all samples are finished.
It then describes the 22 auxiliary functions, including conditioning time, keyboard lock, manual operation of motors and valves, pressure units, method sequence, leak test, temperature calibration and firmware loading, and ends with a procedure for cleaning the headspace tubing with steam from distilled water. This manual is written in Chinese.
| Auxiliary (vial pressurization) gas pressure | 0.68-0.72 bar |
|---|---|
| Number of selectable methods | 9 |
| Number of auxiliary functions | 22 |
| Maximum number of set cycles | 5 |
| Default PRESS times | ON 0 min, OFF 0.2 min |
| Default VENT times | ON 0.2 min, OFF 0.4 min |
| Default SAMPL times | ON 0.4 min, OFF 0.6 min |
| Default probe withdrawal time | 0.65 min |
Press EXTRA FNCT and then the up arrow until EXTRA FUNCTION 7 PRESSURE VALUES appears, press ENTER, press the up arrow again to reach VIAL PRESSURE, and turn the AUX knob at the top right of the instrument to between 0.68 and 0.72 bar.
Nine methods are available; press METHOD, enter the method number with the number keys and press ENTER.
Steam cleaning is recommended: a new vial with distilled water is heated to boiling and the steam is used to clean the tubing, following the temperatures and valve program given in the manual.
The instrument has 22 auxiliary functions, and entering the password 11111 makes functions 14 to 22 visible.