Perfect PT-2091A Programmable Temperature & Humidity Chamber Operating/Maintenance Manual

This product is a programmable temperature and humidity test chamber, mainly used for high and low temperature and humidity simulation environment testing of products.


Perfect PT-2091A Programmable Temperature & Humidity Chamber Operating/Maintenance Manual - cover page
Brand
Perfect
Category
Other products
Document type
User Manual
Language
Chinese
Pages
27
File format
PDF
File size
26.4 MB
Published
08 November, 2023
Updated
08 November, 2023
MD5 checksum
AE46A109EC18D10F2E7779752AA946DE

About this manual

This operating and maintenance manual covers the Perfect PT-2091A programmable temperature and humidity chamber, a microcomputer-controlled environmental test chamber that subjects test products to high/low temperature and humidity simulated environments. The chamber offers a temperature range of -40°C to 150°C and a humidity range of 20–98% RH, with temperature/humidity control precision of ≤±0.3°C and ±2.5% RH.

The manual describes the chamber's system structure, working principle, mechanical construction, temperature/humidity control range, required equipment, installation site and power supply, operating program, maintenance precautions, and troubleshooting examples. The unit uses a Korean-made TEMI-880 touch-screen programmable controller with Chinese/English display, up to 120 program patterns and 1,200 segments, plus an RS-232 communication interface (software sold separately), and a French-imported Tecumseh hermetic compressor with eco-friendly refrigerant. This manual is written in Chinese.

What's inside

  • Performance Specifications
  • System Structure
  • Working Principle
  • Mechanical Structure
  • Temperature and Humidity Control Range
  • Required Equipment
  • Installation Site and Method
  • Power Supply Configuration and Installation
  • Operating Program
  • Maintenance Precautions

Specifications

ModelPT-2091A
Temperature Range-40°C to 150°C
Humidity Range20–98% RH
Temperature/Humidity Control Precision≤±0.3°C; ±2.5% RH
Internal Chamber Dimensions400 × 400 × 500 mm (W×D×H)
External Dimensions600 × 1030 × 1630 mm (W×D×H)
Power SupplyAC 220 V, 50 Hz, 1φ3wire, 4 kW
Heating TimeRoom temperature to 100°C within 30 minutes (no load)
Cooling TimeRoom temperature to -40°C within 60 minutes (no load)
Temperature ControllerKorean TEMI-880 touch-screen programmable controller, Chinese/English display

Frequently asked questions

What temperature and humidity ranges does the PT-2091A cover?

The chamber covers -40°C to 150°C and 20–98% RH; the manual notes to refer to the humidity operating range chart.

Does the PT-2091A support computer communication?

Yes, it has an RS-232 communication interface; the software is sold separately.

What power supply does the PT-2091A require?

AC 220 V, 50 Hz, single-phase 3-wire, 4 kW.