JULABO Immersion Coolers Operating Manual describes the features and characteristics of the product.
JULABO FT200, FT400, FT900 and FD200 immersion coolers are used to cool liquids in bath tanks, Dewar vessels, beakers or other containers, or in closed circuits for the FD200. The manual covers intended use, operator responsibility and safety instructions, disposal, EC conformity, warranty conditions, technical specifications, operating controls, preparations, operating procedures, troubleshooting, and cleaning/repair.
The units are not suitable for direct temperature control of foods, semi-luxury foods and tobacco, or pharmaceutical and medical products. Technical specifications include temperature ranges, cooling capacities, refrigerants, dimensions, weights, ambient temperature limits, mains power connections and current draw.
| FT200 temperature range | -20 to +30 °C |
|---|---|
| FT200 cooling capacity | 0.25 kW at +20 °C, 0.15 kW at 0 °C, 0.04 kW at -30 °C |
| FD200 cooling capacity | 0.22 kW at +20 °C, 0.18 kW at +10 °C |
| FT400 cooling capacity | 0.45 kW at +20 °C, 0.36 kW at +10 °C, 0.14 kW at -20 °C, 0.03 kW at -40 °C |
| FT900 cooling capacity | 0.3 kW at +20 °C, 0.27 kW at +10 °C, 0.2 kW at -40 °C, 0.07 kW at -80 °C |
| FT200/FD200 refrigerant | R134a |
| FT400 refrigerant | R404A |
| FT900 refrigerant | R404A/R23 |
| FT200 dimensions | 18 x 27 x 39 cm |
| FD200 dimensions | 18 x 27 x 39 cm |
| FT400 dimensions | 20 x 30 x 43 cm |
| FT200 weight | 18 kg |
| FD200 weight | 16 kg |
| FT400 weight | 24 kg |
| FT900 weight | 50 kg |
| Ambient temperature | 5 to 35 °C |
They are designed for temperature application to specific fluids in a bath tank, such as Dewar vessels, beakers, or other containers.
No. They are not suitable for direct temperature control of foods, semi-luxury foods and tobacco, or pharmaceutical and medical products.
It is employed to cool liquids in closed circuits, generally installed at the intake of a heating circulator.
The units contain the refrigerants R134a, R404A, and R-23.