A new type of transformer DC resistance tester has been developed by Shanghai Laiyang Electric Technology Co., Ltd. The instrument has the characteristics of small size, light weight and large output current.
The LYZZC-III-50A is a DC resistance tester developed by Shanghai Laiyang Electric Technology for fast measurement of transformer DC resistance. The manual describes it as microcontroller-controlled, with a new power supply design, small size, light weight, high output current, automatic discharge and a discharge indicator. It is used to find manufacturing defects such as winding material, welding, loose connections, missing strands and broken wires, as well as hidden problems after operation.
The manual covers safety measures, performance features, technical specifications, panel layout and test operation. The panel has a 128×32 dot-matrix LCD with Chinese operation, an ammeter, a printer, brightness adjustment, reset, select and confirm keys, V+/V- voltage input terminals and I+/I- current output terminals. Wiring diagrams are provided for direct measurement and for the magnetic assist method for Y(N)-d-11 connection groups; the instrument set includes the main unit, a 50A test lead set, a three-core power cord, fuses and an operation manual. This manual is written in Chinese.
| Output current | 5 A, 10 A, 20 A, 50 A |
|---|---|
| 50 A range | 40 μΩ to 400 mΩ |
| 20 A range | 100 μΩ to 1 Ω |
| Operating temperature | 0 to 40 °C |
| Operating humidity | ≤90%RH, no condensation |
| Power supply | AC220V±10%, 50HZ±1HZ |
| Display | 128×32 dot-matrix LCD, Chinese operation |
It provides 5 A, 10 A, 20 A and 50 A output current settings.
Press reset first and wait until the discharge alarm sound stops before switching tap positions.
Wait until discharge has finished and the alarm sound has stopped before removing the leads.