Laiyang LYZZC-III-50A DC Resistance Tester User Manual

A new type of transformer DC resistance tester has been developed by Shanghai Laiyang Electric Technology Co., Ltd. The instrument has the characteristics of small size, light weight and large output current.


Laiyang LYZZC-III-50A DC Resistance Tester User Manual - cover page
Brand
Laiyang
Category
Test Set
Document type
User Manual
Language
Chinese
Pages
5
File format
PDF
File size
251 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
C53BFBE6681E89C5881DE184D468DEE7

About this manual

The LYZZC-III-50A is a DC resistance tester developed by Shanghai Laiyang Electric Technology for fast measurement of transformer DC resistance. The manual describes it as microcontroller-controlled, with a new power supply design, small size, light weight, high output current, automatic discharge and a discharge indicator. It is used to find manufacturing defects such as winding material, welding, loose connections, missing strands and broken wires, as well as hidden problems after operation.

The manual covers safety measures, performance features, technical specifications, panel layout and test operation. The panel has a 128×32 dot-matrix LCD with Chinese operation, an ammeter, a printer, brightness adjustment, reset, select and confirm keys, V+/V- voltage input terminals and I+/I- current output terminals. Wiring diagrams are provided for direct measurement and for the magnetic assist method for Y(N)-d-11 connection groups; the instrument set includes the main unit, a 50A test lead set, a three-core power cord, fuses and an operation manual. This manual is written in Chinese.

What's inside

  • Overview
  • Safety Measures
  • Performance Features
  • Technical Specifications
  • System Introduction
  • Test and Operation Method
  • Precautions
  • Instrument Set
  • After-Sales Service

Specifications

Output current5 A, 10 A, 20 A, 50 A
50 A range40 μΩ to 400 mΩ
20 A range100 μΩ to 1 Ω
Operating temperature0 to 40 °C
Operating humidity≤90%RH, no condensation
Power supplyAC220V±10%, 50HZ±1HZ
Display128×32 dot-matrix LCD, Chinese operation

Frequently asked questions

What output currents does the LYZZC-III-50A provide?

It provides 5 A, 10 A, 20 A and 50 A output current settings.

What must be done before changing taps on an off-load tap-changing transformer?

Press reset first and wait until the discharge alarm sound stops before switching tap positions.

What should be done before removing the test leads?

Wait until discharge has finished and the alarm sound has stopped before removing the leads.