KTM-4B-25 DI-4B-25 DIGITAL TORQUE CHECKER INSTRUCTION MANUAL
This instruction manual covers the Cedar KTM-4B-25 and DI-4B-25 digital torque checkers and carries the document code OSH-Q08. It begins with 18 numbered safety warnings about the workplace, charging the built-in Ni-Cd battery and handling the instrument, then describes the front panel controls, including the power switch, mode selector, direction switch, zero switch, charging socket, LCD, impact counter and timer setting switches.
A specifications section lists the measuring ranges, accuracy, measurement modes, RS-232C output, power supply and dimensions. The manual then explains how to connect the input and output test connectors, how to choose between the PP, TR and PD measurement modes, how to set the automatic clear time, how data is transmitted, how to charge the battery, and what the one-year warranty and calibration service cover; a troubleshooting table closes the document. This manual is written in Chinese.
| Model | KTM-4B-25; DI-4B-25 |
|---|---|
| Measuring range (N·m) | 0.50 to 25.00 |
| Measuring range (kgf·cm) | 5.0 to 250.0 |
| Measuring range (lbf·in) | 5.0 to 220.0 |
| Accuracy | +/- 0.5% for 500 to 5000 digit readings; +/- 3% for 30 to 499 digit readings |
| Display | 4-digit liquid crystal display |
| Measurement modes | PP peak, TR track, PD peak down |
| Data output | RS232C serial interface, 9600 Baud (optional) |
| Power supply | 6 VDC (Ni-Cd rechargeable battery, 1.2V x 5 cells) |
| Auto power off | Switches off automatically after 5 minutes of non-use |
| Charging time | Less than 8 hours |
| Continuous working time | Approximately 13 hours |
| Battery life | More than 300 recharges (varies with working conditions) |
| Dedicated charger | Input AC120V or AC230V 50/60Hz; output DC7.25V (120mA) |
| Dimensions | 150(W) x 115(D) x 40(H) |
| Weight | Approximately 2.1 kg |
There are three modes: PP (peak), which shows the maximum torque value; TR (track), which shows the real-time torque value; and PD (peak down), which holds the value at the moment the load drops.
The range is 0.50 to 25.00 N·m, which corresponds to 5.0 to 250.0 kgf·cm and 5.0 to 220.0 lbf·in.
Charging takes less than 8 hours, and charging for more than eight hours should be avoided because overcharging can cause explosion, overheating or battery leakage.
Connect the AC charger and press the SYSTEM RESET button on the side; after reset, set the measurement mode switch to TR and press CLEAR so the display shows 0.