Yixiao JB4000 Series Intelligent X-γ Radiation Monitor User Manual

The JB4000 series intelligent x-γ radiation monitor is a special instrument for monitoring the dose rate of x and γ radiation in various radioactive workplaces. The series of instruments have the functions of dose rate, cumulative dose and CPS measurement, which makes the application of the instrument more extensive.


Yixiao JB4000 Series Intelligent X-γ Radiation Monitor User Manual - cover page
Brand
易萧
Category
Monitoring Equipment
Document type
User Manual
Language
Chinese
Pages
11
File format
PDF
File size
252 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
99F5A19DD08CC09A6FCFA0B9FE9E79E1

About this manual

The JB4000 series intelligent X-γ radiation monitor is described in the manual as a dedicated instrument for monitoring X- and γ-ray radiation dose rates in radioactive work places. The document covers the JB4000 and JB4000A models, their features, main technical specifications, precautions, panel layout and operating method.

It explains parameter changes, dose-rate, cumulative-dose and CPS measurement, alarm functions, data query and display backlight, and includes an appendix method for measuring X-ray machines. The manual is in Chinese. This manual is written in Chinese.

What's inside

  • Features and Functions
  • Main Technical Specifications
  • Precautions for Use
  • Structure and Panel Layout
  • Operating Method
  • Parameter Modification
  • Measurement
  • Alarms
  • Data Query
  • Appendix

Specifications

Detectorφ30 × 25 mm NaI(Tl) with photomultiplier tube
Sensitivity1 µSv/h ≥ 350 cps
Energy threshold35 KeV
JB4000 dose-rate range0.01–200.00 µSv/h
JB4000A dose-rate range0.01–1500.00 µSv/h
Cumulative dose range0.00–9999.99 µSv
Energy range48 KeV to 3 MeV
Relative basic error≤ ±10%
Measurement time1, 5, 10, 20, 30 seconds, adjustable
Power consumption≤160 mW, excluding display backlight
Alarm thresholds0.25, 2.5, 10.0 µSv/h

Frequently asked questions

How do I turn on the JB4000 monitor?

With the instrument off, press the POWER key on the panel, then press the RESET key.

What units can be used for dose-rate readings?

Dose-rate readings can be shown in µSv/h, µGy/h or µR/h.

What dose-rate alarm thresholds are available?

The instrument offers three thresholds: 0.25 µSv/h, 2.5 µSv/h and 10.0 µSv/h.

How do I restart cumulative dose measurement?

Press START to clear the previous result and start cumulative dose measurement again from 0.