This document is about the user manual of laminar flow clean benches.
Pachane Linha PCR vertical laminar flow clean bench is intended for Class 100 work under ABNT NBR 13.700 and 15.767 and ISO Class 5 under ISO 14644-1. It has a naval aluminum body with anticorrosive treatment and epoxy paint, an AISI 304 stainless steel work area with brushed finish and removable floor, and 100% air renewal to the work environment.
The bench includes a synthetic class G3 pre-filter and a HEPA class A3 filter, UV and fluorescent lamps, an internal auxiliary outlet, and a needle gas or vacuum valve. Digital versions add a UV lamp hour meter and timer, equipment hour meter, differential pressure manometer, and audiovisual alarm for filter saturation, while analog versions use panel buttons. The manual covers installation, operation, cleaning, maintenance, certification, qualification, and specifications. This manual is written in Portuguese.
| Intended work class | Class 100 (ABNT NBR 13.700 and 15.767); ISO Class 5 (ISO 14644-1) |
|---|---|
| Body construction | Naval aluminum with anticorrosive treatment and epoxy paint |
| Work area | AISI 304 stainless steel, brushed finish, removable floor |
| Air renewal | 100% air renewal to the work environment |
| Inclination | 10 degrees |
| Pre-filter | Synthetic, class G3 |
| HEPA filter class | Class A3 (NBR-6401, EU-13 Eurovent 4/4) |
| HEPA filter efficiency | 99.99% for 0.3 micron particles using DOP |
| Internal illumination | 800 to 1100 lux |
| Auxiliary outlets | 10 A at up to 250 V |
Class 100 work under ABNT NBR 13.700 and 15.767, and ISO Class 5 under ISO 14644-1.
No. The manual states it should not be used to manipulate products of biological risk.
99.99% efficiency for particles of 0.3 micron using DOP.
800 to 1100 lux.