The COD-572 chemical oxygen demand (COD) analyzer uses the potassium dichromate method and the potassium permanganate method to determine the chemical oxygen demand in water. The instrument has the characteristics of wide measuring range, high precision, good stability and simple operation.
This Chinese user manual covers the COD-572 chemical oxygen demand (COD) analyzer from Shanghai Yidian Scientific Instrument Co., Ltd. It explains how to install the instrument and its electrodes, prepare reagents, and perform measurements by coulometric titration using either the potassium dichromate (chromium) method or the potassium permanganate (manganese) method.
The manual also describes the instrument's controls and display, its technical performance, data storage and USB communication, daily maintenance, troubleshooting, and the accessories supplied with the standard set. It is intended for industrial wastewater control analysis and for water-quality COD measurement in environmental monitoring. This manual is written in Chinese.
| Basic error | Not more than 2 mg/L |
|---|---|
| Repeatability | Not more than 2 % |
| Electrolysis current ranges | 10 mA, 20 mA, 40 mA (three ranges) |
| Measurement range | 0.0 to 100.0 mg/L, expandable to 100.0 to 1000 mg/L |
| Ambient temperature | 5 to 40 °C |
| Relative humidity | Not more than 85 % |
| Power supply | AC 220 V (±22 V), 50 Hz (±1 Hz) |
It uses coulometric titration for measurement.
The basic range is 0.0 to 100.0 mg/L, and it can be extended to 100.0 to 1000 mg/L.
It can store 200 COD measurement values each for the manganese method and the chromium method.
It is the amount of oxidant consumed when a water sample is treated with a strong oxidant under specified conditions, generally expressed as oxygen in mg/L.