Beich CH906A/CH906D Analog Battery Tester Operation Manual

This manual introduces the use of CH906A/CH906D analog battery tester of Beiche, including instrument installation, power connection, environmental requirements, instrument startup, etc.


Beich CH906A/CH906D Analog Battery Tester Operation Manual - cover page
Brand
BEICH
Category
Test Set
Document type
User Manual
Language
Chinese
Pages
22
File format
PDF
File size
813 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
D553413F31D2D51924C5B16552E12181

About this manual

The Beich CH906A/CH906D is an analog battery tester that combines charging and discharging in one unit and can simulate a battery's charge and discharge process. It uses a linear power supply for low output ripple and can be used in place of a battery for functional or performance testing of PCBA boards in adapters and power banks; it can also be used as an adjustable power supply.

The manual covers safety warnings, shipment and power checks, fuse installation, power cord connection, environmental requirements, startup, technical specifications, front and rear panel descriptions, display areas, and menu operation. The tester has a 4.3-inch TFT touchscreen, keypad and knob, six voltage/current preset groups, remote sensing, a USB host port, and an RS232C interface. This manual is written in Chinese.

What's inside

  • Checking Shipment
  • Checking Power
  • Installing the Fuse
  • Connecting the Power Cord
  • Environmental Requirements
  • Starting the Instrument
  • Technical Specifications
  • Front Panel Introduction
  • Rear Panel Introduction
  • Menu Operation

Specifications

ModelCH906A / CH906D
Output (discharge) voltageCH906A: 0.1–20 V; CH906D: 0.1–10 V
Output (discharge) currentCH906A: 0–10 A; CH906D: 0–2 A
Input (charge) voltageCH906A: 0.2–20 V; CH906D: 0.2–10 V
Input (charge) currentCH906A: 0–10 A at 0.1–10 V and 0–5 A at 10.01–20 V; CH906D: 0–2 A
Input power110 V/220 V AC
Dimensions230 mm D × 350 mm W × 88 mm H
WeightApprox. 7 kg (CH906A); approx. 4 kg (CH906D)
RippleVoltage: max. 60 mV
Load regulationVoltage: <0.06%

Frequently asked questions

What power source does the instrument require?

It requires 220/110 (1 ± 10%) VAC, 47–63 Hz, with maximum power consumption of 30 VA.

What fuse type is specified?

The manual specifies a 250 V/1 A slow-blow miniature fuse.

What are the normal operating temperature and humidity conditions?

Normal operation should be at 0°C to 40°C with relative humidity of 75% or less.

Does the tester support both charging and discharging?

Yes. It includes battery charging and discharging functions and can switch between them automatically.