This manual provides detailed information on the AQM-ZR safety helmet flame retardancy performance tester, including product features, applicable standards, technical indicators, operation steps and precautions.
This is the Chinese-language instruction manual for the Ulke AQM-ZR safety helmet flame retardancy performance tester (安全帽阻燃性能测试仪), published by Wuhan Ulke Power Equipment Co., Ltd. The manual states the instrument is used to determine the flame retardancy of safety helmets, engineering plastics and helmets that have flame-retardant requirements, and that it can be used by the rubber industry, plastics industry, inspection departments and research units.
The document lists the instrument's features and technical specifications, including an electrically heated wire ignition, an automatically timed and positioned burner, a TedeTal-type burner with 10 mm inner diameter, a 50 mm flame length scale, and a combustible gas leak alarm. It also cites the applicable standard GB/T2812-2006, describes how to clamp a helmet in the specimen holder, gives eight numbered operating steps, and closes with safety precautions covering gas handling, ventilation, protective clothing and grounding. This manual is written in Chinese.
| Burner inner diameter | 10 mm (TedeTal type) |
|---|---|
| Burner moving motor power consumption | 6 W |
| Igniter timer range | 0 to 999.9 seconds, freely settable (10 seconds used internationally) |
| Afterflame timer range and accuracy | 0 to 999.9 seconds, accuracy ±0.01 second |
| Flame length measuring device | scale length 50 mm |
| Gas used | Propane, butane, petroleum liquefied gas |
| Working power supply | AC 220 V; 50 Hz; 50 W |
| Weight | Approximately 30 kg |
| Applicable standard | GB/T2812-2006 Safety Helmet Test Methods (replaces GB/T2812-1989) |
It is stated to follow the national standard GB/T2812-2006, Safety Helmet Test Methods, which replaces GB/T2812-1989.
The manual lists propane, butane and petroleum liquefied gas.
The flame jet head moves to the helmet test area and the burner runs for 10 seconds before the gas is shut off and the burner returns.
It runs on AC 220 V, 50 Hz, with 50 W power consumption.