Kedida CT-6023/CT-6023L Pen-Type pH Meter User Manual

CT-6023/CT-6023L pen type pH meter user manual, including product features, technical parameters, operation guide and other information.


Kedida CT-6023/CT-6023L Pen-Type pH Meter User Manual - cover page
Brand
Kedida
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
3
File format
PDF
File size
403 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
91BADA2ED4115763C124445C57273A0E

About this manual

This user manual covers the Kedida CT-6023 and CT-6023L pen-type pH meters. It describes the pH and temperature LCD display, hold function, waterproof design, automatic temperature compensation, auto power-off, three-point calibration, and the CT-6023L automatic LED backlight.

The manual gives technical parameters and explains factory reset, °C/°F switching, calibration with pH 4.01, 6.86, and 9.18 buffer solutions, over-range and low-battery indications, warranty terms, and how to prepare buffer solutions from the supplied powder packs. This manual is written in Chinese.

What's inside

  • Features
  • Technical parameters
  • Restore factory settings
  • Temperature unit conversion
  • Automatic temperature compensation (ATC)
  • Hold function
  • Calibration
  • Over-range display
  • Low battery indication
  • Buffer solution preparation

Specifications

Measuring range0.00 to 14.00 pH
pH resolution0.01 pH
pH accuracy±0.02 pH
pH calibrationpH 4.01, pH 6.86, pH 9.18 three-point calibration
Temperature display0.1℃
Temperature accuracy1℃/2℉
Temperature compensation0~50℃
Operating temperature0~50℃ (32~122℉)
Dimensions188 x 38 mm including electrode
Weight82 g including electrode
DisplayLCD, 20 mm x 27 mm
Battery1.5 V battery (LR44, BAT) x 4 pcs

Frequently asked questions

How do I calibrate the CT-6023/CT-6023L pH meter?

The manual describes three-point calibration using pH 4.01, pH 6.86, and pH 9.18 buffer solutions, starting with pH 6.86.

What should I do before using the pH meter for the first time?

Soak the electrode in distilled water or pure water for about 10 minutes before first use.

How do I change the temperature unit between ℃ and ℉?

With the meter off, press and hold CAL and ON/OFF together until the ℃ or ℉ symbol appears, then use CAL to select the unit and HOLD to save.

What does “- - -” mean on the pH display?

It appears when the pH value is below 0 or above 14.

Download PDF (403 KB)