CT-6023/CT-6023L pen type pH meter user manual, including product features, technical parameters, operation guide and other information.
This user manual covers the Kedida CT-6023 and CT-6023L pen-type pH meters. It describes the pH and temperature LCD display, hold function, waterproof design, automatic temperature compensation, auto power-off, three-point calibration, and the CT-6023L automatic LED backlight.
The manual gives technical parameters and explains factory reset, °C/°F switching, calibration with pH 4.01, 6.86, and 9.18 buffer solutions, over-range and low-battery indications, warranty terms, and how to prepare buffer solutions from the supplied powder packs. This manual is written in Chinese.
| Measuring range | 0.00 to 14.00 pH |
|---|---|
| pH resolution | 0.01 pH |
| pH accuracy | ±0.02 pH |
| pH calibration | pH 4.01, pH 6.86, pH 9.18 three-point calibration |
| Temperature display | 0.1℃ |
| Temperature accuracy | 1℃/2℉ |
| Temperature compensation | 0~50℃ |
| Operating temperature | 0~50℃ (32~122℉) |
| Dimensions | 188 x 38 mm including electrode |
| Weight | 82 g including electrode |
| Display | LCD, 20 mm x 27 mm |
| Battery | 1.5 V battery (LR44, BAT) x 4 pcs |
The manual describes three-point calibration using pH 4.01, pH 6.86, and pH 9.18 buffer solutions, starting with pH 6.86.
Soak the electrode in distilled water or pure water for about 10 minutes before first use.
With the meter off, press and hold CAL and ON/OFF together until the ℃ or ℉ symbol appears, then use CAL to select the unit and HOLD to save.
It appears when the pH value is below 0 or above 14.