PC-304R9/422R9 PC-R9 series high acceleration life test chamber is a small pressure vessel that can be used for long-term testing under high temperature and high humidity conditions. The maximum pressure can reach 0.393MPa.
The document is a product leaflet for the Hirayama PC-304R9 and PC-422R9 high acceleration life test chambers. It lists both models' external dimensions, test chamber and effective chamber sizes, net weight, power supply, heater capacity, pressure range, temperature and humidity control, test time, safety alarms and standard accessories.
The leaflet describes a programmable temperature and humidity controller with a touchscreen and remote control through external terminals. It covers two control modes (unsaturated steam control for HAST/PCT and PTC saturated control for 100%RH), three exhaust modes, an electronic door lock with sliding tray, bias terminals, and a separate test chamber and steam generator design. This manual is written in Chinese.
| PC-304R9 external dimensions | W710 × D950 × H1620 mm |
|---|---|
| PC-422R9 external dimensions | W710 × D950 × H1620 mm |
| PC-304R9 test chamber dimensions | Ф300 × D522 (34.5 L) |
| PC-422R9 test chamber dimensions | Ф420 × D657 (84.4 L) |
| PC-304R9 net weight | 260 kg |
| PC-422R9 net weight | 295 kg |
| PC-304R9 power supply | 1φ AC220V/230V/240V 50/60Hz |
| PC-422R9 power supply | 1φ AC220V/230V/240V 50/60Hz |
| PC-304R9 heater capacity | 2.3 kW total (steam generator 1.5 kW, test chamber 0.6 kW) |
| PC-422R9 heater capacity | 3.0 kW total (steam generator 1.5 kW, test chamber 1.3 kW) |
| PC-304R9 pressure range | 0.019–0.393 MPa |
| PC-422R9 pressure range | 0.019–0.208 MPa |
| Humidity range | 65–100%RH |
| Humidity control accuracy | ±3%RH (at 85%RH) |
| Heat-up time | Approx. 70 minutes from room temperature to 120℃/85%RH |
| Maximum test time | Up to 500 hours |
| Chamber material | SUS316L stainless steel (electrolytic polishing) |
| Standard bias terminals | 20 per chamber (125 V, 30 W total) |
The humidity range is 65–100%RH.
Twenty bias terminals are provided per chamber as standard.
It takes approximately 70 minutes from room temperature to reach 120℃/85%RH.
The test time can be set up to 500 hours.