This manual details the wiring, parameter settings, display and operation, product performance, alarm information, troubleshooting and other contents of the L-magBP series electromagnetic flowmeter converter, which is an essential reference for using this product.
This document is the user manual for the L-magBP series electromagnetic flowmeter converter, featuring a round dial and ordinary LCD display. It details the installation and wiring procedures, including signal lines, excitation lines, and power connections. The guide covers various parameter settings such as flow units, alarm thresholds, output configurations (current, pulse, frequency), and sensor parameters. This manual is written in Chinese.
| Maximum Counter Value | 999999999 |
|---|---|
| Measurement Damping Time Range | 1S to 60S |
| Current Output Range | 4-20mA |
| Pulse Width Range | 0.5 to 1999ms |
| Frequency Output Range | 1 to 5000 Hz |
| Communication Modes | MODBUS, HART, PROFIBUS |
| Communication Speed Range | 300 to 38400 |
| Pipe Diameter Range | 3 to 3000 |
The instrument supports units such as L/s, L/m, L/h, m3/s, m3/m, m3/h, uk/s, uk/m, uk/h, us/s, us/m, us/h, kg/s, kg/m, kg/h, t/s, t/m, and t/h.
The excitation current line should use a two-core insulated rubber soft cable, with a recommended model of RVVP2*0.3mm2.
The converter supports MODBUS, HART, and PROFIBUS communication modes.