Scribner Associates 840/850/855 Humidifiers Accelerated Cooling Application Note

The document describes how to rapidly cool the humidifier in a fuel cell test system.


Scribner Associates 840/850/855 Humidifiers Accelerated Cooling Application Note - cover page
Brand
Scribner Associates
Category
Other products
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
112 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
7FAC626F28D8C874D182D859D510EB3C

About this manual

This application note from Scribner Associates Inc. (rev. A, 11/2017) describes a manual method for accelerated cooling of the humidifiers in Scribner fuel cell test systems. Because the systems are designed for rapid heating but not rapid cooling, natural cooling from around 80°C can take hours. The note explains how sequential draining and filling of the humidifiers with cold pressurized water cools them from 80°C to about 30°C in 20–30 minutes.

The document lists the required tools (a manual ¼ inch Swagelok 3-way valve, T-fitting, tubing and wrenches), the test conditions such as 310 kPa water pressure and 100 kPa back pressure, and the drain/fill timing needed to avoid triggering the Low Water Alarm. It also includes a setup figure and example cooling curves for the anode and cathode humidifiers.

What's inside

  • Introduction
  • Tools
  • Test Conditions

Specifications

Initial humidifier temperature80 °C
Target final humidifier temperature30 °C
Back pressure100 kPa (~15 PSI) on anode and cathode
Flow rate1 SLM
Water pressure310 kPa (45 PSI)
Auto water fill cycle13 pulses, ~1 sec each, 20 sec between pulses (~250 sec total)

Frequently asked questions

How long does natural cooling of the humidifier take?

Natural cooling of the humidifiers from typical elevated temperature testing (e.g., ~80°C) can take hours.

How can the Low Water Alarm be avoided during rapid cooling?

Manually fill the humidifier until the water is about 1 cm from the top of the sight glass before the 4-minute auto fill cycle ends, so the low water level sensor is satisfied at the end of the cycle.

How long does the accelerated cooling procedure take?

Two drain/auto water fill cycles take about 10 minutes per humidifier, and the total process took just under 30 minutes in the example shown.

What tools are required for the manual cooling method?

A manual ¼ inch Swagelok 3-way valve, a ¼ inch Swagelok T-fitting, ¼ inch tubing, and a 9/16 inch or adjustable wrench.

Download PDF (112 KB)