The document describes how to rapidly cool the humidifier in a fuel cell test system.
This application note from Scribner Associates Inc. (rev. A, 11/2017) describes a manual method for accelerated cooling of the humidifiers in Scribner fuel cell test systems. Because the systems are designed for rapid heating but not rapid cooling, natural cooling from around 80°C can take hours. The note explains how sequential draining and filling of the humidifiers with cold pressurized water cools them from 80°C to about 30°C in 20–30 minutes.
The document lists the required tools (a manual ¼ inch Swagelok 3-way valve, T-fitting, tubing and wrenches), the test conditions such as 310 kPa water pressure and 100 kPa back pressure, and the drain/fill timing needed to avoid triggering the Low Water Alarm. It also includes a setup figure and example cooling curves for the anode and cathode humidifiers.
| Initial humidifier temperature | 80 °C |
|---|---|
| Target final humidifier temperature | 30 °C |
| Back pressure | 100 kPa (~15 PSI) on anode and cathode |
| Flow rate | 1 SLM |
| Water pressure | 310 kPa (45 PSI) |
| Auto water fill cycle | 13 pulses, ~1 sec each, 20 sec between pulses (~250 sec total) |
Natural cooling of the humidifiers from typical elevated temperature testing (e.g., ~80°C) can take hours.
Manually fill the humidifier until the water is about 1 cm from the top of the sight glass before the 4-minute auto fill cycle ends, so the low water level sensor is satisfied at the end of the cycle.
Two drain/auto water fill cycles take about 10 minutes per humidifier, and the total process took just under 30 minutes in the example shown.
A manual ¼ inch Swagelok 3-way valve, a ¼ inch Swagelok T-fitting, ¼ inch tubing, and a 9/16 inch or adjustable wrench.