Nuova Fima MN15 MN16 MN17 MN25 MN17L Test and Laboratory Pressure Gauges User Guide

This document is the user guide of Nuova Fima MN15-16-17-17L-25 pressure gauges, which introduces the application, packaging, installation, accuracy, zeroing, and usage range of pressure gauges.


Nuova Fima MN15 MN16 MN17 MN25 MN17L Test and Laboratory Pressure Gauges User Guide - cover page
Brand
Nuova Fima
Category
Other
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
207 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
312385D2439FD7D34E21443A96D885B5

About this manual

Nuova Fima's user guide covers the MN15, MN16, MN17, MN25 and MN17L test and laboratory pressure gauges. It states that the instruments are designed for use as test instruments to check or recalibrate other instruments and are produced in conformity with EN 837-1-2 and ASME B40.1.

The document explains packing, installation, accuracy and calibration reports, zeroing, limits of use for ambient and process temperature and working pressure, wrong applications, maintenance and disposal. It notes that calibration and mounting positions different from standard are shown on the dial.

What's inside

  • General information
  • Intended use
  • Packing
  • Installation
  • Accuracy
  • Zeroing
  • Limits of use
  • Working pressure
  • Maintenance
  • Disposal

Specifications

MN15 ambient / process temperature-25…+65°C / -40…150°C
MN16 ambient / process temperature-25…+65°C / -40…150°C
MN17 ambient / process temperature+15…+65°C / max 65°C
MN25 ambient / process temperature-20…+65°C / max 65°C
MN17/L ambient / process temperature+10…+60°C / max 20°C
AccuracyWithin ±3°C of the ambient temperature shown on the dial
Recommended working pressureBetween 25% and 75% of full scale range
Zero correction limit5% of full scale value

Frequently asked questions

What are the Nuova Fima MN15/MN16/MN17/MN25/MN17L gauges intended for?

They are designed as test instruments to check or recalibrate other instruments where accuracy and repeatability are important.

How do I correct zero deviation on the MN17/L?

Use the external adjusting knob on the front protection ring.

What is the recommended working pressure range?

The instrument should be selected with a working pressure between 25% and 75% of the full scale range.

What is recommended if zero deviation is larger than 5% of full scale?

The manual recommends proceeding with the full recalibration procedure.

Download PDF (207 KB)