Nikko Hansen TN100IR Portable Turbidimeter Instruction Manual

This manual introduces the use, calibration and maintenance methods of TN100IR turbidimeter.


Nikko Hansen TN100IR Portable Turbidimeter Instruction Manual - cover page
Brand
Nikko Hansen
Category
Other products
Document type
User Manual
Language
Japanese
Pages
12
File format
PDF
File size
695 KB
Published
02 November, 2023
Updated
02 November, 2023
MD5 checksum
11CDFF56DB7B46E0A7D68B67FD682A5F

About this manual

The TN100IR is a compact portable turbidimeter based on scattered-light (nephelometric) measurement in accordance with ISO 7027 and DIN 27027. The manual explains package contents, part names, keypad functions, battery installation, reset, calibration, measurement, troubleshooting, maintenance, and accessories.

Calibration uses four standard solutions at 0.02, 20.0, 100, and 800 NTU. The manual also describes sample vial preparation, measurement precautions, automatic power-off after about 20 minutes of inactivity, and handling and storage of the instrument and vials. This manual is written in Japanese.

What's inside

  • Package contents
  • Part names and keypad functions
  • Battery installation and replacement
  • Resetting the instrument
  • Calibration vial preparation
  • Calibration procedure
  • Measurement precautions and sample vial preparation
  • Measurement procedure
  • Troubleshooting
  • Product specifications and accessories

Specifications

Measurement methodScattered light (nephelometry), ISO 7027 and DIN 27027 compliant
Calibration standards0.02, 20.0, 100, and 800 NTU
Batteries4 AAA batteries
Auto power offAbout 20 minutes of inactivity
Calibration error limit±10% of factory NTU initial setting (0.08 NTU or more for 0.02 NTU)

Frequently asked questions

What batteries does the TN100IR use?

It uses four AAA batteries.

When does the TN100IR turn off automatically?

It turns off automatically after about 20 minutes without operation.

Which calibration standards are supplied with the TN100IR?

The supplied calibration set includes 0.02, 20.0, 100, and 800 NTU solutions.

Is an error display a fault if no vial is installed at power-on?

No, an error display appears but it is not a malfunction.