This manual introduces the use, calibration and maintenance methods of TN100IR turbidimeter.
The TN100IR is a compact portable turbidimeter based on scattered-light (nephelometric) measurement in accordance with ISO 7027 and DIN 27027. The manual explains package contents, part names, keypad functions, battery installation, reset, calibration, measurement, troubleshooting, maintenance, and accessories.
Calibration uses four standard solutions at 0.02, 20.0, 100, and 800 NTU. The manual also describes sample vial preparation, measurement precautions, automatic power-off after about 20 minutes of inactivity, and handling and storage of the instrument and vials. This manual is written in Japanese.
| Measurement method | Scattered light (nephelometry), ISO 7027 and DIN 27027 compliant |
|---|---|
| Calibration standards | 0.02, 20.0, 100, and 800 NTU |
| Batteries | 4 AAA batteries |
| Auto power off | About 20 minutes of inactivity |
| Calibration error limit | ±10% of factory NTU initial setting (0.08 NTU or more for 0.02 NTU) |
It uses four AAA batteries.
It turns off automatically after about 20 minutes without operation.
The supplied calibration set includes 0.02, 20.0, 100, and 800 NTU solutions.
No, an error display appears but it is not a malfunction.