This document is about SentronicD Digital Electronic Pressure Regulator Series 608/609 with Display and Keypad installation manual
ASCO NUMATICS SentronicD Series 608/609 is a digital electronic pressure regulator with display and keypad. This installation manual covers description, catalogue numbering, operating elements, manual pressure regulation, operating modes, electrical and pneumatic connections, factory settings, field-programmable settings, technical characteristics, accessories, maintenance, and dimensions and weights.
The manual includes electrical connection diagrams and pin assignments for the 5-pin M12 connector, analog setpoint and outlet pressure relationships, pressure range limits, and factory PID control parameters. It also describes shut-off, overtemperature, undervoltage/overvoltage, and autosafe operating modes.
| Nominal diameter / flow rate (DN 4) | DN 4 mm, 470 l/min (SRA) |
|---|---|
| Nominal diameter / flow rate (DN 8) | DN 8 mm, 1300 l/min (SRA) |
| Maximum outlet pressure | 3, 6 or 10 bar |
| Maximum inlet pressure | 6, 9 or 13 bar |
| Supply voltage | 24 V DC, +15%/-10% tolerance, max. 10% ripple |
| Digital output | PNP, max. 200 mA / 4.8 W |
| Analog setpoint input | 0-10 V, 0-20 mA or 4-20 mA |
| Analog feedback output | 0-10 V, 0-20 mA or 4-20 mA |
| Pneumatic connections | G 1/8, G 1/4, G 3/8, NPT 1/8, NPT 1/4 or NPT 3/8 |
| Cable | 5-wire 2 m; 6-wire 5 m or 10 m |
| Connector | 5-pin M12 female |
| Overtemperature threshold | 100 °C |
| Undervoltage / overvoltage | below 20 V / above 30 V |
| Shut-off threshold | below 0.5% setpoint |
| Autosafe current limit | 1000 mA (DN 8) or 560 mA (DN 4) for more than 20 s |
| Factory setpoint | 0 bar at 0 V / 0 mA / 4 mA |
| Factory span | 3, 6 or 10 bar at 10 V / 20 mA |
| Controller | PID; proportional gain 4.0; integration 0.1 s; derivation 8 ms |
Press both arrow buttons beneath the display during power-up; the display shows H n d and the yellow LED stays on.
The coil current is switched off and the valve is fully exhausted.
24 V DC with a tolerance of +15%/-10% and a maximum ripple of 10%.
It is off in normal operation, on for low voltage, and flashing for overvoltage.