This manual is the instruction manual of BilgMon488 bilge alarm, which introduces the product's construction, operation, precautions, specifications, installation, calibration, operating instructions and other contents
The BilgMon488 is a 15 ppm bilge alarm from Brännströms Elektronik, designed for use with 15 ppm oil-water separator units. It has two adjustable alarms triggered when the oil content of the processed sample exceeds the set limit, with relay and LED outputs plus a 0(4)-20 mA current output corresponding to 0-30 ppm. The unit consists of a MASTER unit (housing with LCD, buttons and LEDs) and a hermetically sealed SENSOR unit that communicates wirelessly with the MASTER. The operating log can be downloaded through a USB interface.
This instruction manual covers construction, operation and cautions, specifications for the 115/230 V AC and 24 V AC/DC models, installation, calibration, operating instructions, simulation, contrast, automatic stopping device test, response test, calibration check at 40 NTU, log entries, maintenance, spare parts and troubleshooting.
| Oil range | 0 – 30 ppm |
|---|---|
| Resolution | 0.1 ppm |
| Response time | < 3 sec |
| Alarm points 1 and 2 | 1-15 ppm, user adjustable |
| Alarm 1 (valve control) delay | 0-10 sec, user adjustable |
| Alarm 2 (annunciation) delay | 0-600 sec, user adjustable |
| Alarm hysteresis | 0.5 ppm (below alarm point) |
| Current output | 0 – 20 mA or 4 – 20 mA for 0 – 30 ppm |
| Supply (115/230 V AC model) | 1 A, 115 or 230 V AC, 50 – 60 Hz |
| Supply (24 V AC/DC model) | 1.5 A, 24 V AC (50 – 60 Hz) or 24 V DC |
| LCD display | 2x16 alphanumeric display |
| Control | 4 button keypad |
It is designed specifically for use in conjunction with 15 ppm oil-water separator units.
No. Opening the SENSOR unit invalidates the calibration, so it shall not be opened.
The operating log can be downloaded through a USB interface, which requires driver software and cables supplied on demand.
Use a soft bottle-brush and mild detergent, and avoid anything that could scratch the glass tube inside the SENSOR unit.