This manual introduces the use of FTI 3000
This manual provides detailed instructions for operating the FTI 3000 software component burner, including configuration parameters, cabling options, and diagnostic tools. It covers setup for diode testing, power supply configuration, and troubleshooting with failure indicators. The guide also explains how to use the BIAG diagnostics application and manage calibration settings.
The document includes sections on operation tabs, phase meter, light pole status, and system health monitoring. It details how to configure diode strings, manage heating currents, and interpret failure symbols. Users can also access quick keys and diagnostic tabs for advanced testing and calibration tasks.
| Vf Accuracy | +/- 0.5% of measured value |
|---|---|
| Vz Accuracy | +/- 0.8% of measured value |
| IHeat Accuracy | +/-10% of measured value |
| Meter Accuracy (100mV Range) | 0.15% of Value +/- 1mV |
| Maximum Operating Condition | Maximum 600V. 2.5A per IHeat current source. |
Tj °C is calculated using the formula: Tj °C = 25 + ((Vf@25°C – Vf measurement) / Kfactor).
Use the 'Clear Flash Data' button to clear existing data, or the 'Erase Flash Memory' button to completely wipe the flash memory.