CD97 Hi-Tech TDS Meter with ATC User Manual
The Milwaukee Instruments CD97 Hi-Tech TDS Meter with ATC measures total dissolved solids in solution. Operation involves removing the protective cap, turning the meter on, immersing it up to the maximum immersion level, stirring gently until the display stabilizes, and reading the value, with automatic temperature compensation from 5 to 50ºC.
Calibration uses the optional M10080B solution at 800 ppm by adjusting the calibration trimmer with a small screwdriver until the reading matches the solution value. The meter uses four 1.5V alkaline batteries with an approximate life of 300 hours, and the instrument is warranted for 2 years from purchase, with the probe warranted for 6 months.
| Range | 0 to 999 ppm |
|---|---|
| Resolution | 1 ppm |
| Accuracy | ±10 ppm at 20°C |
| Temperature Compensation | Automatic from 5 to 50ºC |
| Environment | 5 to 50°C, 95 % RH max. |
| Battery Type | 4 x 1.5V alkaline |
| Battery Life | approximately 300 hours of use |
| Dimensions | 150 x 30 x 24 mm |
| Weight | 85 g |
The range is 0 to 999 ppm.
Immerse the tester in the M10080B calibration solution and adjust the trimmer until the reading matches 800 ppm.
Pull out the battery compartment and replace all four 1.5V batteries, paying attention to their polarity.
This instrument is warranted for 2 years from the date of purchase.