Haitongda MLN Series Marsh Funnel Viscometer User Manual

Martens-type funnel viscosity meter is a simple device for measuring viscosity. It is manufactured according to the American API standard, and the viscosity of drilling fluid is measured by the time of a fixed amount of drilling fluid flowing out of the funnel.


Haitongda MLN Series Marsh Funnel Viscometer User Manual - cover page
Brand
Haitongda
Category
Other
Document type
User Manual
Language
Chinese
Pages
9
File format
PDF
File size
424 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
0A11F9EAD8F9596CD83D8A51A533224B

About this manual

This Chinese user manual describes the Haitongda MLN series Marsh funnel viscometer, a simple device for the routine measurement of viscosity. According to the manual, it is built to the American API standard and determines the viscosity of drilling fluid by measuring the time a fixed volume of fluid takes to flow out of the funnel. Six configurations are listed: MLN-2, MLN-3, MLN-3A, MLN-3E, MLN-4 and MLN-4A, which differ in funnel and receiving-cup material (plastic or stainless steel) and in whether a stopwatch is included.

The manual covers the instrument's structure and working principle, its technical parameters, calibration and performance check with distilled water, the test procedure for drilling fluid samples, maintenance and cleaning of the funnel, measuring cup and receiving cup, transport and storage, a fault-diagnosis table, the list of supplied accessories and the one-year warranty terms. It also includes important safety notes and the manufacturer's contact details. This manual is written in Chinese.

What's inside

  • Overview
  • Specifications and Models
  • Technical Parameters
  • Structure and Working Principle
  • Operating Instructions
  • Maintenance and Care
  • Transport and Storage
  • Fault Diagnosis and Troubleshooting
  • Supplied Accessories and Tools List
  • Warranty

Specifications

Working environment temperature5℃ to 35℃
Screen mesh aperture1.68 mm (12 mesh)
Funnel volume1500 ± 15 mL
AccuracyOutflow time of 946 mL distilled water at 21℃±3℃ is 26 s ± 0.5 s
Transport and storage standardGB/T 25480-2010
Warranty periodOne year

Frequently asked questions

What outflow time should the viscometer show when it is checked with distilled water?

At 21℃±3℃, 946 mL of distilled water should flow out in 26 s ± 0.5 s.

Which models are available?

The manual lists MLN-2, MLN-3, MLN-3A, MLN-3E, MLN-4 and MLN-4A; they differ in funnel and cup material and whether a stopwatch is supplied.

What causes a flow rate greater than the standard flow rate?

A worn, oversized inner diameter of the flow tube; the manual says to replace the funnel.

How long is the warranty?

The warranty period is one year from the production date shown on the product test sheet.