This manual introduces the basic operation procedures and usage methods of YEGK-C12 high voltage switch characteristic testing instrument, aiming to help users to use the product correctly, safely and efficiently.
The YEGK-C12 high voltage switch characteristic testing instrument measures mechanical characteristic parameters of high-voltage circuit breakers, including SF6, GIS, vacuum, oil and special breakers. It is designed with reference to GB1984-2003 and DL/T846.3-2004.
In one breaker operation it can measure 12 metal contact break channels plus 6 main and 6 auxiliary channels, recording time, bounce, speed, coil current and resistance. Features include a 7-inch color screen, internal adjustable DC power supply, storage for 100 test groups, USB and RS232 interfaces, and a high-speed thermal printer. This manual is written in Chinese.
| Break channels | 12 metal contact, 6 main, 6 auxiliary |
|---|---|
| Time measurement range | Internal trigger 0.01-10000 ms; external trigger 0.01-20000 ms |
| Time accuracy | Within 200 ms: 0.05% ± 1 digit |
| Speed measurement range | 1 mm sensor 0.01-25.00 m/s; 0.1 mm sensor 0.001-2.50 m/s; 345° sensor 1 wave/0.01° |
| Stroke measurement range | Linear sensor 0-50 mm (resolution 0.1 mm); optional 100-1000 mm; rotary sensor 0-1000 mm (resolution 0.1°) |
| Current and resistance range | Max current 30 A (resolution 0.01 A); resistance 1000 Ω (resolution 0.01 Ω) |
| Instrument power supply | AC 220 V ±10%, 50 Hz ±10% |
| DC power output | DC 20-270 V adjustable; DC 220 V ≤20 A (short time); 4400 W |
| External trigger voltage | AC/DC 10-300 V, current ≤120 A |
| Host dimensions | 360 × 260 × 170 mm |
| Operating environment | -20 to +50 °C |
| Relative humidity | ≤90% |
| Storage capacity | 100 test groups |
| Display | 7-inch color high-brightness screen |
| Interfaces | USB and RS232 |
| Printer | High-speed thermal printer |
| Interference immunity | Internal anti-interference circuit for use in 500 kV substations |
The instrument requires AC 220 V ±10% at 50 Hz ±10%.
It can measure 12 metal contact break channels, 6 main channels and 6 auxiliary channels.
It provides an adjustable DC 20-270 V output, up to 20 A at DC 220 V for short periods, with 4400 W power.
The host can store 100 groups of test data.