The HYG2670F insulation and dielectric strength tester is a high-performance insulation and dielectric strength tester. It can simultaneously display insulation resistance, dielectric strength and breakdown current values, and has multiple test functions. The high voltage output accuracy of this series of testers is high (better than 5%), the leakage current test accuracy is high (better than 5%), and the insulation resistance test speed is fast. It is suitable for rapid flow testing on production lines and for laboratory multi-function high precision requirements. The measurement principle of the instrument conforms to the safety performance test requirements of GB4706.1, GB3883.1 and GB4943, and can be used to test the electrical strength and insulation resistance of devices with basic insulation, additional insulation and reinforced insulation. It is suitable for safety testing of household appliances, electric power equipment, communication equipment, automobiles, machinery and other industries.


Huaneng Yangguang HYG2670F Insulation and Dielectric Strength Tester User Manual - cover page
Brand
Huaneng Yangguang
Category
Test Set
Document type
User Manual
Language
Chinese
Pages
24
File format
PDF
File size
421 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
495C73FE0B23BBB72E80756F44888169

About this manual

The HYG2670F is an AC withstand voltage and insulation resistance tester designed for electrical safety testing of appliances, instruments, office equipment, and power tools. It provides simultaneous display of insulation resistance, test voltage, and breakdown current with LED digital readouts.

The manual details safety rules, installation requirements, front and rear panel functions, operating procedures for insulation and withstand voltage tests, technical specifications including voltage and current ranges, troubleshooting guidance, and calibration procedures for both withstand voltage and insulation resistance measurements. This manual is written in Chinese.

What's inside

  • Safety Rules
  • Installation Points
  • Product Overview
  • Front Panel Description
  • Rear Panel Description
  • Accessories
  • Technical Specifications
  • Troubleshooting
  • Quality Assurance
  • Calibration Procedures

Specifications

AC Withstand Voltage Output Range0.3–5 kV ±5%
Output Frequency Accuracy50 Hz ±0.5%
Output WaveformSine wave, crest factor 1.3–1.5
Insulation Resistance Ranges20 MΩ, 200 MΩ, 2000 MΩ
Insulation Test VoltagesDC 500 V, DC 1000 V
Breakdown Current Ranges2 mA, 20 mA, 100 mA
DisplayThree LED digital displays for insulation resistance, voltage, and current
Operating Environment0–40 °C, 20–80% RH
Input Voltage220 Vac ±10%, 50/60 Hz
Fuse5 A / 8 A, 250 V slow-blow
Dimensions280 mm (W) × 89 mm (H) × 380 mm (D)

Frequently asked questions

What is the input voltage requirement for the HYG2670F?

The tester requires 220 Vac ±10% at 50/60 Hz.

What insulation resistance measurement ranges are available?

The instrument offers three ranges: 20 MΩ, 200 MΩ, and 2000 MΩ.

Which test voltages can be selected for insulation resistance testing?

DC 500 V and DC 1000 V are selectable for insulation resistance measurement.

Does the HYG2670F support DC withstand voltage testing?

No, the HYG2670F model does not have the AC/DC selection function for withstand voltage testing.